Hemolysis of red blood cells after cell washing with different automated technologies: clinical implications in a neonatal cardiac surgery population.
O'Leary MF, Szklarski P, Klein TM, Young PP
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of the method of washing RBCs and post-wash storage duration on levels of extracellular K and hemolysis. This study used group A+ non-irradiated leukoreduced RBCs in additive solution 3 (AS-3) that had been refrigerated for 4 days. An aliquot was centrifuged immediately before washing, and 0, 4, 6, 12 and 24 h after washing, and the supernatant was stored at -20 degrees C until analysis. The interim storage temperature was unspecified.
Summary of Findings
Washing caused an immediate decline in extracellular potassium concentrations, regardless of the washing machine used. Generally, after the initial washing, the extracellular K increased during post-wash refrigerated storage, with the largest increases observed when specimens were washed using the COBE model 1, and almost no change in extracellular K when specimens were washed with the Fresnus (p<0.001 at 12 h). Similarly, the hemolysis index increased with washing and during post-wash refrigerated storage, with the largest increases observed when specimens were washed using the COBE model 1, and significantly smaller increases observed in specimens washed with the Fresnus (p<0.001 at 4 h). Specimens washed using the COBE model 2 displayed increases in hemolysis index and extracellular K that were intermediate to the other two washing platforms. Importantly, by 24 h post-wash with the COBE model 1, the extracellular K was higher than the pre-wash levels.
Biospecimens
CellRed Blood CellsPreservation Types
FrozenDiagnoses
NormalPlatforms
Cell count/volumeSpectrophotometryElectrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood processing method COBE 2991 model 1COBE 2991 model 2Fresnus continuous autotransfusion systemUnwashedStorage Storage duration 0 h12 h24 h4 h6 hAnalyte Extraction and Purification Washing COBE 2991 model 1COBE 2991 model 2Fresnus continuous autotransfusion systemUnwashed