Effects of EDTA and Sodium Citrate on hormone measurements by fluorometric (FIA) and immunofluorometric (IFMA) methods.
Kohek M, Leme C, Nakamura IT, De Oliveira SA, Lando V, Mendonca BB
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of anticoagulant type and the use of serum versus plasma on the measurement of 15 hormone analytes. The influence of analysis by FIA versus IFMA assays was also discussed. Serum was collected in serum separator tubes (SST).
Summary of Findings
The levels of LH, FSH, PRL, and free T4 were not significantly different between EDTA plasma, sodium citrate plasma, and serum. However, SHBG, GH, TSH, INS, and CPEP levels (measured with IFMA assays) were lower in EDTA plasma than in serum, and levels of total T3, total T4, E2, testosterone, cortisol, and progesterone (measured with FIA assays) were higher in EDTA plasma than in serum. Further, TSH and INS levels (measured with IFMA assays) were lower in sodium citrate plasma than in serum, and total T3, total T4, E2, testosterone, and progesterone levels (measured with FIA assays) were higher in sodium citrate plasma than in serum. The authors did not offer a direct comparison of hormone levels between EDTA and sodium citrate plasma.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
GlycoproteinClinical chemistry/auto analyzerPeptideClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant NonePotassium EDTASodium citrateBiospecimen Aliquots and Components Blood and blood products PlasmaSerumClinical chemistry/auto analyzer Specific Technology platform Fluorometric assayImmunofluorometric assay