Washing of banked blood by three different blood salvage devices.

Gruber M, Breu A, Frauendorf M, Seyfried T, Hansen E

Publication: Transfusion (2013) • Vol 53 • Page 1001-9PMID: 22897672View on PubMed

Purpose

The purpose of this paper was to determine the effects of washing red blood cells (RBCs) with different autotransfusion devices on RBC quality after refrigerated storage.

Conclusion

The RBC recovery, and iomeperol, citrate and potassium removal were highest from undiluted RBCs when the Electa device was used for washing, and the hematocrit, protein and free hemoglobin (fHb) removal were highest from undiluted RBCs when the OrthoPAT device was used for washing. Dilution of the RBCs in saline increased RBC recovery and the elimination of potassium, citrate, iomeperol (spiked in), protein and fHb. pH was not affected by washing.

Studies (1)

Purpose

The purpose of this study was to determine the effects of washing RBCs with different autotransfusion devices on RBC quality after refrigerated storage. Out-dated RBCs (40-43 days old) in citrate-phosphate-dextrose (CPD) saline-adenine-glucose-mannitol (SAGM) were filtered and left undiluted, or diluted 1:2 or 1:4 in saline before loading into the different autotransfusion devices for washing.

Summary of Findings

The mean RBC recovery was highest at 94.5% using the Electa device for washing, regardless of predilution, and was lowest at 52.0% from undiluted specimens with the OrthPat. Mean RBC recovery using the OrthoPat device for washing was increased to 86.8% when specimens were diluted 1:2 (p=0.001) and to 92.8% when specimens were diluted 1:4 (p=0.001). When the CATS device was used for washing, RBC recovery increased non-significantly upon dilution from 80.8% (undiluted specimens) to 84.5% (specimens diluted 1:2) and 93.7% (specimens diluted 1:4). After washing, the hematocrit increased from 32.3% to 72.6% for specimens washed with the OrthoPAT (p<0.001), but lesser increases of 59.1% for the CATS, and 54.8% for the Electa were observed compared to the OrthoPAT (p<0.001, both), and there was no effect of pre-dilution. More than 90% of the spiked in iomeprol was removed using the Electa, with or without dilution, and the same was true in diluted RBCs when the CATS or OrthoPAT were used for washing. Importantly, with a 1:4 dilution, the Electa, OrthoPAT and CATS were able to remove 99.3%, 98.5%, and 99.1% of the spiked in iomeprol, respectively. Similarly, elimination of citrate and potassium from undiluted specimens was highest using the Electa followed by the OrthoPAT, but after 1:4 dilution, all three machines performed citrate and potassium elimination similarly. Washing improved removal of protein and fHb, but OrthoPAT performed slightly better than the Electa using undiluted specimens. The CATS autotransfuser was least effective at removal of fHb and protein from undiluted specimens. pH was not affected by washing.

  • Biospecimens

    CellRed Blood Cells
  • Preservation Types

    Other Preservative
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalIon selective electrode
    ProteinClinical chemistry/auto analyzer
    Small moleculeHPLC
    Small moleculepH
    Cell count/volumeHematology/ auto analyzer
    Cell count/volumeFlow cytometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    RBCs diluted 1:2 with saline
    RBCs diluted 1:4 with saline
    Undiluted RBCs
    Biospecimen Aliquots and ComponentsBlood processing method
    Washed with CATS
    Washed with Electa
    Washed with OrthoPAT
    Analyte Extraction and PurificationWashing
    CATS
    Electa
    OrthoPAT