Washing of banked blood by three different blood salvage devices.
Gruber M, Breu A, Frauendorf M, Seyfried T, Hansen E
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of washing RBCs with different autotransfusion devices on RBC quality after refrigerated storage. Out-dated RBCs (40-43 days old) in citrate-phosphate-dextrose (CPD) saline-adenine-glucose-mannitol (SAGM) were filtered and left undiluted, or diluted 1:2 or 1:4 in saline before loading into the different autotransfusion devices for washing.
Summary of Findings
The mean RBC recovery was highest at 94.5% using the Electa device for washing, regardless of predilution, and was lowest at 52.0% from undiluted specimens with the OrthPat. Mean RBC recovery using the OrthoPat device for washing was increased to 86.8% when specimens were diluted 1:2 (p=0.001) and to 92.8% when specimens were diluted 1:4 (p=0.001). When the CATS device was used for washing, RBC recovery increased non-significantly upon dilution from 80.8% (undiluted specimens) to 84.5% (specimens diluted 1:2) and 93.7% (specimens diluted 1:4). After washing, the hematocrit increased from 32.3% to 72.6% for specimens washed with the OrthoPAT (p<0.001), but lesser increases of 59.1% for the CATS, and 54.8% for the Electa were observed compared to the OrthoPAT (p<0.001, both), and there was no effect of pre-dilution. More than 90% of the spiked in iomeprol was removed using the Electa, with or without dilution, and the same was true in diluted RBCs when the CATS or OrthoPAT were used for washing. Importantly, with a 1:4 dilution, the Electa, OrthoPAT and CATS were able to remove 99.3%, 98.5%, and 99.1% of the spiked in iomeprol, respectively. Similarly, elimination of citrate and potassium from undiluted specimens was highest using the Electa followed by the OrthoPAT, but after 1:4 dilution, all three machines performed citrate and potassium elimination similarly. Washing improved removal of protein and fHb, but OrthoPAT performed slightly better than the Electa using undiluted specimens. The CATS autotransfuser was least effective at removal of fHb and protein from undiluted specimens. pH was not affected by washing.
Biospecimens
CellRed Blood CellsPreservation Types
Other PreservativeDiagnoses
NormalPlatforms
Electrolyte/MetalIon selective electrodeProteinClinical chemistry/auto analyzerSmall moleculeHPLCSmall moleculepHCell count/volumeHematology/ auto analyzerCell count/volumeFlow cytometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components RBCs diluted 1:2 with salineRBCs diluted 1:4 with salineUndiluted RBCsBiospecimen Aliquots and Components Blood processing method Washed with CATSWashed with ElectaWashed with OrthoPATAnalyte Extraction and Purification Washing CATSElectaOrthoPAT