Skip to main content

Effect of anticoagulants and storage temperatures on stability of plasma and serum hormones.

Evans MJ, Livesey JH, Ellis MJ, Yandle TG

Publication: Clin Biochem (2001) • Vol 34 • Page 107-12PMID: 11311219View on PubMed

Purpose

The purpose of this paper was to determine the effects of anticoagulant type, storage duration, and storage temperature on a group of 22 hormones in plasma and serum.

Conclusion

Storage of specimens at -20 degrees C until analysis led to anticoagulant-specific differences for a number of hormone analytes. In general, hormone analytes were far less stable when stored at 30 degrees C before being frozen than they were when stored at 4 degrees C. The authors conclude that fluoride and EDTA served as better anticoagulants for hormone analysis in specimens requiring storage and transport at 4 degrees C for 120 h than heparin or the use of serum. Adrenocorticotrophic hormone (ACTH) was notably the least stable analyte at 4 degrees C. While stability was low across the board when specimens were stored at 30 degrees C, in general, analytes were slightly more stable in EDTA plasma than when the other anticoagulants or serum were used, but analyte-specific exceptions were observed. Brain natriuretic peptide (BNP) and N-terminal BNP (NT-BNP) showed analyte-, anticoagulant-, and blood component-specific stabilities ranging from 4 to 20 h when either whole blood or plasma was stored at room temperature.

Studies (2)

Purpose

The purpose of this study was to determine the effects of anticoagulant type, storage duration, and storage temperature on a group of 22 hormones in plasma and serum. After experimental storage, all specimens were frozen at -20 degrees C until analysis. Actual measurements were only taken after 24 or 120 h of storage of specimens at 4 or 30 degrees C, but the time elapsing before the median immunoreactivity changed by 10% or more was interpolated (duration of stability).

Summary of Findings

When specimens were stored at -20 degrees C until analysis (no initial storage at 4 or 30 degrees C), 6 of 22 analytes had significantly different levels between serum and EDTA plasma, 4 of 20 analytes had significantly different levels between fluorinated plasma and EDTA plasma, and 3 of 22 had significantly different levels between hepariniated plasma and EDTA plasma. In general, hormone analytes were far less stable when stored at 30 degrees C before being frozen than they were when stored at 4 degrees C. Only 5 of 19 analytes showed anticoagulant-specific durations of stability at 4 degrees C that were observed before the 120 h time point . The authors conclude that fluoride and EDTA served as better anticoagulants for hormone analysis in specimens requiring storage and transport at 4 degrees C for 120 h than heparin or the use of serum. ACTH was notably the least stable analyte at 4 degrees C. On the other hand, only alpha-subunit total, C-peptide, and follicle stimulating hormone (FSH) were stable for >120 h at 30 degrees C for all anticoagulants and serum. While stability was low across the board when specimens were stored at 30 degrees C, in general, analytes were slightly more stable in EDTA plasma than when the other anticoagulants or serum were used, but analyte-specific exceptions were observed.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    PeptideImmunoradiometric assay
    GlycoproteinRadioimmunoassay
    PeptideRadioimmunoassay
    SteroidRadioimmunoassay
    ProteinRadioimmunoassay
    GlycoproteinImmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    EDTA
    Lithium heparin
    Sodium fluoride/potassium oxalate
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    StorageStorage duration
    0 h
    120 h
    24 h
    StorageStorage temperature
    -20 degrees C
    30 degrees C
    4 degrees C