Skip to main content

Effects of contamination of blood specimens with liquid potassium-EDTA anticoagulant.

Davidson DF

Publication: Ann Clin Biochem (2002) • Vol 39 • Page 273-80PMID: 12038602View on PubMed

Purpose

The purpose of this paper was to determine the effects of refrigerated storage and the addition of EDTA to serum on 24 clinical chemistry analytes.

Conclusion

Refrigerated storage of serum for 72 h significantly affected 9 of 24 analytes, while the addition of EDTA to serum significantly affected 11 of 24 analytes.

Studies (1)

Purpose

The purpose of this study was to determine the effects of refrigerated storage and the addition of EDTA to serum on 24 clinical chemistry analytes. EDTA-free lyophilized quality control serum was reconstituted with various concentrations of EDTA. Spiked specimens were analyzed 20 minutes after the addition of EDTA (assumed to be held at room temperature), and then again after 72 h at 4 degrees C.

Summary of Findings

Without the addition of EDTA, 72 h of refrigerated storage of serum led to significant increases in cholesterol and urea and decreases in HDL-cholesterol, urate, bilirubin, creatine kinase (CK), aspartate transaminase (AST), alanine transaminase (ALT), and lactate dehydrogenase (LDH). On the other hand, 72 hours of refrigerated storage alone had no significant effects on albumin, chloride, gamma-glutamyltransferase (GGT), iron, phosphate, sodium, total protein, triglycerides, calcium, magnesium, unsaturated iron-binding capacity (UIBC), bicarbonate, amylase activity, or alkaline phosphatase (ALP). The addition of 5.0 mmol/L EDTA to serum resulted in significant decreases in bicarbonate and ALP, both at the 20 min time point and after 72 h of refrigerated storage, a complete mitigation of the decreases in AST and ALT after 72 h of refrigerated storage, a partial mitigation of the decreases in CK and LDH after 72 h of refrigerated storage, significant decreases in amylase activity, calcium, magnesium, and UIBC, and a significant increase in potassium. AST, ALT, CK, LDH, and amylase activity were not significantly affected by the addition of EDTA at the 20 min time point. The authors show that the effects of EDTA addition on measurement of UIBC, calcium, magnesium, and ALP increased with increasing EDTA concentration. The addition of EDTA had no effects on cholesterol, HDL-cholesterol, urate, urea, bilirubin, albumin, chloride, GGT, iron, phosphate, sodium, total protein, or triglycerides.
  • Biospecimens

    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    0 mmol/L EDTA
    0.5 mmol/L EDTA
    1.0 mmol/L EDTA
    1.5 mmol/L EDTA
    2.0 mmol/L EDTA
    2.5 mmol/L EDTA
    3.0 mmol/L EDTA
    3.5 mmol/L EDTA
    4.0 mmol/L EDTA
    4.5 mmol/L EDTA
    5.0 mmol/L EDTA
    StorageStorage duration
    20 min
    72 h