Skip to main content

Hemoglobin by first derivative spectrophotometry: extent of hemolysis in plasma and serum collected in vacuum container devices.

Copeland BE, Dyer PJ, Pesce AJ

Publication: Ann Clin Lab Sci (1989) • Vol 19 • Page 383-8PMID: 2802517View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type, anticoagulant type, and collection volume on the amount of hemoglobin in serum and plasma specimens, as well as to determine the effects of storing serum on the clot.

Conclusion

On average, hemoglobin levels were significantly lower in plasma than in serum. Citrated plasma had significantly lower amounts of hemoglobin than heparinated or EDTA plasma. There were no significant differences between hemoglobin levels in serum stored in the original tube on the clot and serum transferred to new tubes for storage, or between serum collected in plain tubes and that collected in serum separator tubes (SST). Further, the size of the collection tube, and consequently, the specimen volume, did not affect hemoglobin levels among filled 5, 7, and 10 mL SST or 15 mL plain serum tubes.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube and anticoagulant type, and collection volume on the amount of hemoglobin in serum and plasma specimens, as well as to determine the effects of storage of serum on the clot. Heparinated and citrated plasma and serum collected in plain tubes but not intended for storage on the clot were transferred to new plain tubes and stored at 4 degrees C for 1 to 7 days prior to analysis, but serum collected in SST was stored in the original container, and EDTA plasma was held at room temperature for 24-30 hours prior to analysis.

Summary of Findings

On average, hemoglobin levels were significantly lower in plasma than in serum. Citrated plasma had significantly lower amounts of hemoglobin than heparinated or EDTA plasma. There were no significant differences between hemoglobin levels in serum stored in the original tube on the clot and serum transferred to new tubes for storage, or between serum collected in plain tubes and that collected in SST. Further, the size of the collection tube (and collection volume) did not affect hemoglobin levels among filled 5, 7, and 10 mL SST or 15 mL plain serum tubes.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)Other Preservative
  • Diagnoses

    Not specified
  • Platforms

    ProteinSpectrophotometry
    ProteinCoulter counter
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Citrate
    EDTA
    Heparin
    None
    Biospecimen AcquisitionType of collection container/solution
    Plain serum tube
    SST
    Biospecimen Aliquots and ComponentsAliquot size/volume
    10 mL
    15 mL
    5 mL
    7 mL
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen PreservationType of fixation/preservation
    None (fresh)
    Refrigeration
    StorageType of storage container
    New plain serum tube
    Original plain serum tube