Preanalytical storage does not affect 99th percentile cardiac troponin T concentrations measured with a high-sensitivity assay.
Gillis JM, Dunselman P, Jarausch J, de Jong N, Cobbaert CM
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of delayed blood processing and storage of plasma and serum on levels of cTnT1 in blood from patients with suspected myocardial damage. Blood was collected into Venosafe serum tubes or lithium heparin gel tubes. Blood was processed immediately or after storage at room temperature for 3 or 5 h, and the plasma and serum were frozen. Additional plasma and serum specimens were stored at room temperature for 1, 2, 6 or 24 h, 2-8 degrees C for 1, 2, 6 or 24 h, or -20 degrees C for 24 h, 1 week, 4 weeks, or 3 months, after which all specimens were frozen at -80 degrees C for 3 months.
Summary of Findings
cTnT1 levels in serum or lithium heparin plasma were not affected by storage of whole blood for up to 5 h before processing (longest duration examined). Further, there was no effect of storage of plasma or serum at room temperature, 2-8 degrees C, or -20 degrees C on cTnT1 levels. cTnT1 levels were comparable in lithium heparin plasma and serum.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
FrozenDiagnoses
Cardiovascular DiseaseNot specifiedPlatforms
PeptideClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products PlasmaSerumWhole bloodBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage duration 0 h1 h1 week2 h24 h3 h3 months4 weeks5 h6 hStorage Storage temperature -20 degrees C-80 degrees C2-8 degrees CRoom temperature