Scaling of hemolysis in needles and catheters.

Sharp MK, Mohammad SF

Publication: Ann Biomed Eng (1998) • Vol 26 • Page 788-97PMID: 9779951View on PubMed

Purpose

The purpose of this paper was to determine the effects of cannula gauge, cannula material, and pressure at the upstream end of the cannula on hemolysis among plasma specimens.

Conclusion

Hemolysis increased with increasing pressure inside the cannula and with increasing cannula diameter (lower gauge). At low pressures at the upstream end of the needle or catheter, there were no significant differences in hemolysis due to cannula material (stainless steel, Teflon, or polyurethane). At higher pressures, differences due to cannula material were variable between cannula gauges.

Studies (1)

Purpose

The purpose of this study was to determine the effects of cannula gauge, cannula material, and pressure at the upstream end of the cannula on hemolysis among plasma specimens. To test the effects of cannula gauge and material, blood specimens inside a syringe surrounded by a pressure chamber were transferred into test tubes through stainless steel needles or catheters of different gauges.

Summary of Findings

Hemolysis increased with increasing pressure inside the cannula and with increasing cannula diameter (lower gauge), regardless of cannula material. At low pressures at the upstream end of the needle or catheter, there were no significant differences in hemolysis due to cannula material (stainless steel, Teflon, or polyurethane). At higher pressures, passing blood through 22 gauge Teflon catheters, as opposed to stainless steel or polyurethane, tended to result in lower rates of hemolysis. In addition, at higher pressures, passing blood through 18 gauge stainless steel needles, as opposed to teflon or polyurethane, resulted in higher rates of hemolysis. There were no statistically significant differences in hemolysis rates, with regard to cannula material, for 14 gauge cannulas.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    ProteinSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionCannulation
    119 kPa pressure
    14 gauge
    171 kPa pressure
    18 gauge
    18.2 kPa pressure
    207 kPa pressure
    22 gauge
    25.2 kPa pressure
    261 kPa pressure
    316 kPa pressure
    391 kPa pressure
    512 kPa pressure
    Polyurethane
    Stainless steel
    Teflon
    Biospecimen AcquisitionMethod of fluid acquisition
    IV catheter
    Needle
    Biospecimen AcquisitionNeedle gauge
    14 gauge
    18 gauge
    22 gauge