Preanalytical aspects and sample quality assessment in metabolomics studies of human blood.

Yin P, Peter A, Franken H, Zhao X, Neukamm SS, Rosenbaum L, Lucio M, Zell A, Häring HU, Xu G, Lehmann R

Publication: Clin Chem (2013) • Vol 59 • Page 833-45PMID: 23386698View on PubMed

Purpose

The purpose of this paper was to determine if the use and type of anticoagulant, the degree of hemolysis, a room temperature delay in processing, and freeze-thaw cycling significantly affect the metabolite profile of serum and plasma.

Conclusion

The degree of chemical noise observed during liquid chromatography mass spectrometry (LC-MS) was influenced by anticoagulant, with the least amount of noise present in specimens collected into tubes containing either potassium EDTA or sodium fluoride. Hemolysis resulted in 69 features that differed significantly compared to control specimens (p<0.05). A room temperature delay in processing resulted in 64 features, which displayed robust increases in peak area after 2 or 4 h at room temperature. Such storage effects were absent when specimens were stored in ice water for 2 or 4 h instead of at room temperature. Subjecting plasma specimens to 2 or more freeze-thaw cycles (thawed in ice water) elicited a significant difference in 4 metabolites.

Studies (4)

Purpose

The purpose of this study was to assess potential impacts of anticoagulant use and type on chemical noise in during mass spectroscopy analyses. S-monovette collection tubes containing 4 different anticoagulants (lithium heparinate, sodium fluoride, sodium citrate, potassium EDTA) or kaolin (serum) were used for blood collection. Blood specimens from three volunteers were evaluated by mass spectroscopy for chemical noise [(-CH2CH2O-)n]. Controls were completed for each tube type investigated and contained saline with 50 g/L human albumin. Plasma specimens were gently mixed and stored in ice water for 25 min, while serum specimens were permitted to clot for 25 min at room temperature prior to centrifugation.

Summary of Findings

Chemical noise inference was clearly detected as mass spectroscopy signals in blood specimens collected in tubes containing plastic beads, which included lithium heparinate and serum collection tubes. Collection tubes with the least amount chemical noise compared to control tubes were potassium EDTA and sodium fluoride collection tubes.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Small moleculeLC-MS or LC-MS/MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Lithium heparin
    None
    Potassium EDTA
    Sodium citrate
    Sodium fluoride