Discard first tube for coagulation testing.
Tekkesin N, Esen OB, Kilinç C, Eviyaoglu O
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of aliquot sequential collection on PT and APTT using specimens from patients on oral anticoagulant therapy. 2 specimens were consecutively drawn into tubes containing sodium citrate.
Summary of Findings
Significant differences were found in PT, APTT, and INR between plasma from the first collection and plasma from the second collection. The differences in PT between collection tubes ranged from -15.4% to 18.2%. The differences in APTT between collection tubes ranged from -16.86% to 27.99%. And lastly, the differences in INR between first and second collection tubes ranged from -28.6% to 23.7%, and the difference in INR between the two tubes increased with increasing INR values.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level INR 2.0-3.0INR 3.1-4.5INR <2INR >4.5Biospecimen Aliquots and Components Aliquot sequential collection 1st collection2nd collection