Stability of haematology parameters on the LH750: comparison of Sarstedt Monovettes and BD Vacutainers.
Sethi B, Taylor EC, Taylor J, Mitchell P
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type and refrigerated storage duration on complete blood counts in EDTA-anticoagulated blood.
Summary of Findings
MPV increased significantly with refrigerated storage of blood in vacutainers (p=0.043) and monovette tubes (p=0.026) and exceeded the recommended precision limits in specimens refrigerated for 18 h or more in vacutainer tubes or 24 h or more in monovette tubes. WBC counts exceeded the precision limits in specimens refrigerated in either tube type for 48 h or more. Neutrophil counts increased slightly with storage in monovette tubes (p=0.0034), but not vacutainer tubes (p=0.0871), and consequently, neutrophil counts exceeded the precision limits in specimens refrigerated in monovette tubes for 48 h. Although all other analytes remained within the precision limits for the 72 h of refrigerated storage, monocyte (MO) counts and basophil (BA) counts were significantly affected by storage in vacutainers (p=0.0136 and p=0.0025, respectively) and monovette tubes (p=0.0174 and p=0.0206, respectively), but the counts were higher than baseline at some time-points and lower at others.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Other PreservativeDiagnoses
NormalPlatforms
Cell count/volumeHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution MonovetteVacutainerBiospecimen Preservation Type of fixation/preservation None (fresh)RefrigerationStorage Storage duration 12 h24 h30 h36 h4 h42 h48 h60 h72 h8 h<30 min