Preanalytical evaluation of serum 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 measurements using LC-MS/MS.

Mochizuki A, Kodera Y, Saito T, Satoh M, Sogawa K, Nishimura M, Seimiya M, Kubota M, Nomura F

Publication: Clin Chim Acta (2013) • Vol 420 • Page 114-20PMID: 23123828View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type, anticoagulant type, delayed blood separation, storage, freeze-thaw cycling, patient food consumption, and analysis method on the measurement of 25-hydroxyvitamin D3 (25OHD3) in plasma and serum from healthy patients.

Conclusion

Collection tube type, anticoagulant type, storage of uncentrifuged blood, plasma, or serum, and freeze-thaw cycling had no effects on the measurement of 25OHD3 in plasma and serum. A non-significant trend to lower 25OHD3 in specimens collected 1 h after eating a standardized meal compared to levels in specimens collected before patients ate or 2-3 h after they ate were noted. 25OHD3 levels measured by LC-MS/MS and RIA were very strongly correlated. Average 25OHD3 levels were higher in specimens from Japanese males than in specimens from Japanese females. Interference of the LC-MS/MS assay by bilirubin C, bilirubin F, hemolytic hemoglobin, chyle, and immunoglobulin M rheumatoid factor (IgM-RF) was less than 10%.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube type, anticoagulant type, delayed blood separation, storage, freeze-thaw cycling, patient food consumption, and analysis method on the measurement of 25OHD3 in plasma and serum from healthy patients. Plasma and serum were stored at -80 degrees C until analysis.

Summary of Findings

Neither storage of blood for up to 6 h at room temperature prior to centrifugation nor storage of plasma or serum impacted the measurement of 25OHD3 in serum or plasma, regardless of light exposure during storage. Further, 25OHD3 levels in serum and plasma were not affected by up to 5 freeze-thaw cycles. While no significant effects were found based on the time of specimen collection, a trend toward lower 25OHD3 levels was noted in specimens collected 1 h after patients ate a standardized meal compared to levels in specimens collected before eating or 2 h after eating. Collection tube and anticoagulant type had no effects on 25OHD3 measurements. 25OHD3 levels measured by LC-MS/MS and RIA were very strongly correlated (r=0.947) with RIA giving a positive bias of 4 ng/mL. Average 25OHD3 levels were higher in Japanese males than in females (25.5 ng/mL and 20.9 ng/mL, respectively). Interference of the LC-MS/MS assay by bilirubin C, bilirubin F, hemolytic hemoglobin, chyle, and IgM-RF was less than 10%.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    SteroidLC-MS or LC-MS/MS
    SteroidRadioimmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPatient gender
    Female
    Male
    Biospecimen AcquisitionAnticoagulant
    None
    Potassium EDTA
    Sodium EDTA
    Sodium fluoride
    Sodium heparin
    Biospecimen AcquisitionTime of biospecimen collection
    1 h after eating
    2 h after eating
    3 h after eating
    Before eating
    Biospecimen AcquisitionType of collection container/solution
    Insepack plasma tubes with dipotassium EDTA
    Insepack plasma tubes with dipotassium EDTA and sodium fluoride
    Insepack plasma tubes with disodium EDTA
    Insepack plasma tubes with sodium heparin
    Insepack serum tubes for rapid coagulation
    Insepack serum tubes with separating gel and silica
    Microtainer plasma tubes for neonates with dipotassium EDTA
    Microtainer serum tubes for neonates (gel and thrombin, accelerating clotting agent as the additive)
    Serum clot activator
    Serum separator (SST)
    VENOjectII plain serum tube
    Biospecimen Aliquots and ComponentsBiospecimen components
    Spiked with IgM-RF
    Spiked with bilirubin C
    Spiked with bilirubin F
    Spiked with chyle
    Spiked with hemolytic hemoglobin
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Whole blood
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageFreeze/thaw cycling
    1 cycle
    2 cycles
    3 cycles
    4 cycles
    5 cycles
    StorageStorage conditions
    Covered tube
    Uncovered tube
    StorageTime at room temperature
    1 h
    15 min
    2 h
    30 min
    4 h
    6 h
    LC-MS or LC-MS/MS SpecificTechnology platform
    RIA