Sodium citrate vacuum tubes validation: preventing preanalytical variability in routine coagulation testing.

Lima-Oliveira G, Lippi G, Salvagno GL, Montagnana M, Picheth G, Guidi GC

Publication: Blood Coagul Fibrinolysis (2013) • Vol 24 • Page 252-5PMID: 23160241View on PubMed

Purpose

The purpose of this paper was to determine the effects of vacuum tube brand on activated partial thromboplastin time (APTT), prothrombin time (PT), and fibrinogen measurements in sodium citrate plasma.

Conclusion

APTT and PT measurements were significantly different between specimens collected in the majority of tube brands examined. However, fibrinogen measurements were not significantly affected by the brand of vacuum tube used for blood collection.

Studies (1)

Purpose

The purpose of this study was to determine the effects of vacuum tube brand on APTT, PT, and fibrinogen measurements in sodium citrate plasma. Plasma was frozen at -70 degrees C until analysis.

Summary of Findings

APTT measurements were statistically and clinically significantly different between specimens collected in the Venosafe tubes and the Vacuette tubes and between specimens collected in the Labor Import tubes and the Venosafe, Vacuette, Vacutainer, and Monovette tubes. Specimens collected in the Vacuette tubes yielded the shortest average APTT while specimens collected in the Labor Import tubes yielded the longest average APTT. Statistically significant but clinically insignificant differences in APTT measurements were also seen between specimens collected in the Venosafe and Vacutainer tubes. PT measurements were statistically and clinically significantly different between specimens collected in the Venosafe tubes and the Vacuette, Vacutainer, and Monovette tubes, between specimens collected in the Vacuette tubes and the Labor Import and Monovette tubes, and between specimens collected in the Labor Import tubes and the Vacutainer and Monovette tubes. Specimens collected in Venosafe and Labor Import tubes yielded the longest average PT, while specimens in the Vacuette tubes yielded the shortest average PT. Statistically significant but clinically insignificant differences in PT measurements were also seen between specimens collected in the Vacutainer tubes and the Vacuette and Monovette tubes. Fibrinogen measurements were not significantly different between specimens collected in any of the tube brands.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    MorphologyHematology/ auto analyzer
    GlycoproteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    BD Vacutainer
    Labor Import
    S-Monovette
    Vacuette
    Venosafe