Adrenocorticotropic hormone stability in preanalytical phase depends on temperature and proteolytic enzyme inhibitor.

Casati M, Cappellani A, Perlangeli V, Ippolito S, Pittalis S, Romano R, Limonta G, Carati L

Publication: Clin Chem Lab Med (2013) • Vol 51 • Page e45-7PMID: 23023695View on PubMed

Purpose

The purpose of this paper was to determine the effects of storing blood at room temperature or in ice water in collection tubes containing aprotinin rather than storing blood in standard tubes in ice water on the measurement of adrenocorticotrophic hormone (ACTH) in K3EDTA plasma.

Conclusion

ACTH concentrations in K3EDTA plasma declined and hemolysis increased with increasing storage duration of blood in ice water, but storage at room temperature rather than 0 degrees C and the addition of aprotinin attenuated these effects such that the effect on hemolysis almost disappeared when blood was stored at room temperature in tubes containing aprotinin.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storing blood at room temperature or in ice water in collection tubes containing aprotinin rather than storing blood in standard tubes in ice water on the measurement of ACTH in K3EDTA plasma. Plasma was stored at -80 degrees C, prior to analysis.

Summary of Findings

ACTH concentrations in K3EDTA plasma declined and hemolysis increased with increasing storage duration of blood in ice water, but storage at room temperature rather than 0 degrees C and the addition of aprotinin attenuated these effects. When blood was stored for 3 or 4 h at room temperature in tubes containing aprotinin, significantly more ACTH was present in the K3EDTA plasma than when blood was stored in tubes without aprotinin in ice water for 3 or 4 h (p=0.01, and p=0.0002, respectively). Significantly more hemolysis was observed when specimens were stored in standard tubes in ice water for 2, 3 or 4 h compared to levels measured in plasma from unstored specimens (p<0.0001, all), but only slightly higher levels of hemolysis were observed when whole blood was stored at room temperature in tubes containing aprotinin compared to levels in plasma from unstored blood.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    PeptideImmunoassay
    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    0 h
    1 h
    2 h
    3 h
    4 h
    StorageStorage temperature
    Room temperature
    Ice water
    Analyte Extraction and PurificationProtease inhibitor
    Aprotinin
    No protease inhibitor added