Sample collection and platelet function testing: influence of vacuum or aspiration principle on PFA-100 test results.

Lippi G, Ippolito L, Zobbi V, Sandei F, Favaloro EJ

Publication: Blood Coagul Fibrinolysis (2013) • Vol 24(6) • Page 666-9PMID: 23470650View on PubMed

Purpose

The purpose of this paper was to determine the effects of manual versus vacuum aspiration of blood, collection tube type, and platelet activation solution on closure time using the PFA-100.

Conclusion

Use of manual aspiration during collection rather than vacuum aspiration resulted in shorter PFA-100 closure times. There were no significant differences between specimens collected by vacuum aspiration into Vacutainer tubes versus Monovettes.

Studies (1)

Purpose

The purpose of this study was to determine the effects of manual versus vacuum aspiration of blood, collection tube type, and platelet activation solution on closure time using the PFA-100. A Vacutainer tube was filled using vacuum aspiration with venipuncture of the right arm, followed by a Monovette tube filled by vacuum aspiration from a second venipuncture on the opposite arm, and lastly, another Monovette tube was filled by manual aspiration from the same second venipuncture site. All specimens were collected into sodium citrate with a tourniquet times of <30 sec.

Summary of Findings

The closure times obtained when blood was collected by manual aspiration into Monovette tubes (SD-A) were significantly shorter than those obtained when blood was collected by vacuum aspiration into either Vacutainer (BD-V) or Monovette tubes (SD-V), regardless of which activation solution was used in the assay. SD-A closure time had a mean bias of -12 s when either collagen and ADP or collagen and epinephrine were used compared to results for SD-V specimens. SD-A closure times had mean biases of -9 s and -14 s when collagen and ADP or collagen and epinephrine were used, respectively, compared to results for BD-V specimens. The differences between SD-V and BD-V specimen closure times were not significant for either activation solution.

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnatomical location of blood draw
    Arm
    Left and right sides compared
    Biospecimen AcquisitionMethod of fluid acquisition
    Syringe draw
    Vacuum tube
    Biospecimen AcquisitionType of collection container/solution
    BD Vacutainer
    Sarstedt S-Monovette
    Hematology/ auto analyzer SpecificReaction solution
    Collagen and ADP
    Collagen and epinephrine