Influence of mechanical hemolysis of blood on two D-dimer immunoassays.
Lippi G, Avanzini P, Zobbi V, Ippolito L
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of mechanically induced hemolysis and assay type on D-dimer measurement in plasma. Hemolysis was induced in aliquots A, B, and C by aspiration of whole blood 1, 2, or 3 times, respectively, through a 30 gauge needle. Hemolysis was not induced in a 4th aliquot. 4 pools of blood were used for analysis including a range of D-dimer concentrations.
Summary of Findings
As expected, increasing the number of passages of whole blood through the fine needle induced higher levels of hemolysis. The concentrations of D-dimer in plasma decreased as the concentration of cell free hemoglobin increased, regardless of the type of immunoassay. The decreases in plasma D-dimer concentrations between the non-hemolyzed aliquots and hemolyzed aliquot A were not significant when measured by the HemosIL AcuStar D-dimer immunoassay, but they were significant when measured by the HemosIL D-dimer HS for ACL TOP immunoassay. All other decreases between the non-hemolyzed aliquots and hemolyzed aliquots B and C were significant for both immunoassays. The HemosIL AcuStar D-dimer immunoassay measured significantly higher concentrations of D-dimer in all specimens compared to the HemosIL D-dimer HS for ACL TOP immunoassay.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinSpectrophotometryPeptideImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level High D-dimer concentrationLow D-dimer concentrationMedium D-dimer concentrationVery high D-dimer concentrationBiospecimen Aliquots and Components Hemolysis Fine needle aspiration-inducedNot inducedImmunoassay Specific Technology platform HemosIL AcuStar D-dimerHemosIL D-dimer HS for ACL TOP