Influence of mechanical hemolysis of blood on two D-dimer immunoassays.

Lippi G, Avanzini P, Zobbi V, Ippolito L

Publication: Blood Coagul Fibrinolysis (2012) • Vol 23 • Page 461-3PMID: 22576291View on PubMed

Purpose

The purpose of this paper was to determine the effects of mechanically induced hemolysis and assay type on D-dimer measurement in plasma.

Conclusion

The concentrations of D-dimer in plasma decreased as the concentration of cell free hemoglobin increased, regardless of the type of immunoassay. The decreases in plasma D-dimer concentrations between the non-hemolyzed aliquots and the aliquot hemolzyed by aspiration through a needle once were not significant when measured by the HemosIL AcuStar D-dimer immunoassay, but were significant when measured with the other assay. All other decreases between the non-hemolyzed aliquots and aliquots that were hemolyzed by aspiration through a needle 2 or 3 times were significant for both immunoassays. The HemosIL AcuStar D-dimer immunoassay measured significantly higher concentrations of D-dimer in all specimens compared to the HemosIL D-dimer HS for ACL TOP immunoassay.

Studies (1)

Purpose

The purpose of this study was to determine the effects of mechanically induced hemolysis and assay type on D-dimer measurement in plasma. Hemolysis was induced in aliquots A, B, and C by aspiration of whole blood 1, 2, or 3 times, respectively, through a 30 gauge needle. Hemolysis was not induced in a 4th aliquot. 4 pools of blood were used for analysis including a range of D-dimer concentrations.

Summary of Findings

As expected, increasing the number of passages of whole blood through the fine needle induced higher levels of hemolysis. The concentrations of D-dimer in plasma decreased as the concentration of cell free hemoglobin increased, regardless of the type of immunoassay. The decreases in plasma D-dimer concentrations between the non-hemolyzed aliquots and hemolyzed aliquot A were not significant when measured by the HemosIL AcuStar D-dimer immunoassay, but they were significant when measured by the HemosIL D-dimer HS for ACL TOP immunoassay. All other decreases between the non-hemolyzed aliquots and hemolyzed aliquots B and C were significant for both immunoassays. The HemosIL AcuStar D-dimer immunoassay measured significantly higher concentrations of D-dimer in all specimens compared to the HemosIL D-dimer HS for ACL TOP immunoassay.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    ProteinSpectrophotometry
    PeptideImmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    High D-dimer concentration
    Low D-dimer concentration
    Medium D-dimer concentration
    Very high D-dimer concentration
    Biospecimen Aliquots and ComponentsHemolysis
    Fine needle aspiration-induced
    Not induced
    Immunoassay SpecificTechnology platform
    HemosIL AcuStar D-dimer
    HemosIL D-dimer HS for ACL TOP