Vitamin C in mouse and human red blood cells: an HPLC assay.

Li H, Tu H, Wang Y, Levine M

Publication: Anal Biochem (2012) • Vol 426 • Page 109-17PMID: 22522059View on PubMed

Purpose

The purpose of this paper was to determine the effects of specimen volume and storage of whole blood, red blood cells (RBCs) or processed specimens on levels of ascorbic acid measured by high performance liquid chromatography (HPLC) with coulometric detection.

Conclusion

Ascorbic acid was stable during storage of RBCs at 4 degrees C for up to 24 h, but levels declined with storage of RBCs at 25 degrees C or when whole blood was stored at 4 degrees C. Ascorbic acid was stable in processed RBCs in methanol/EDTA stored at -80 degrees C for 6 weeks.

Studies (1)

Purpose

The purpose of this study was to determine the effects of specimen volume and storage of whole blood, RBCs or processed specimens on levels of ascorbic acid measured by HPLC with coulometric detection. Whole blood was collected into lithium heparin tubes, while packed RBCs were collected into citrate phosphate double dextrose (CP2D) additive solution 3 (AS-3).

Summary of Findings

The use of HPLC with coulometric detection allowed for detection of ascorbic acid at concentrations as low as 26 nM and from as little as 20 uL whole blood or 5 uL RBCs. Ascorbic acid was stable in RBCs stored at 4 degrees C for up to 24 h. In contrast, ascorbic acid levels declined when RBCs were stored at 25 degrees C for as short as 2 h in specimens with a 26 uM concentration of ascorbic acid, but for specimens with a 34 uM concentration of ascorbic acid, the decline was smaller and first apparent after 24 h of storage at 25 degrees C. Ascorbic acid levels declined by an average of 10% when whole blood was stored for 1 week at 4 degrees C. Further, after 6 weeks of storage at 4 degrees C, levels declined by an average of 50%. Ascorbic acid was stable in processed RBCs (in methanol/EDTA) stored at -80 degrees C for 6 weeks. There was a linear relationship between RBC and plasma ascorbic acid levels.

  • Biospecimens

    FluidBlood
    FluidPlasma
    CellRed Blood Cells
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Small moleculeHPLC
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    26 uM ascorbic acid
    34 uM ascorbic acid
    Biospecimen Aliquots and ComponentsAliquot size/volume
    10 uL
    15 uL
    2 uL
    20 uL
    5 uL
    50 uL
    70 uL
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Red blood cells
    Whole blood
    StorageStorage duration
    0 h
    1 h
    1 week
    2 h
    2 weeks
    24 h
    3 weeks
    4 h
    4 weeks
    5 weeks
    6 h
    6 weeks
    StorageStorage temperature
    25 degrees C
    4 degrees C