Variation in clinical vitamin D status by DiaSorin Liaison and LC-MS/MS in the presence of elevated 25-OH vitamin D2.
de Koning L, Al-Turkmani MR, Berg AH, Shkreta A, Law T, Kellogg MD
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of measuring 25-OH vitamin D levels in plasma by the DiaSorin Liaison assay versus by LC-MS/MS in specimens containing high and low levels of 25-OH vitamin D2. Plasma was stored at 4 degrees C for less than 7 days prior to analysis.
Summary of Findings
Significantly lower 25-OH vitamin D concentrations were measured by the Liaison analyzer than by LC-MS/MS (average of 26 nmol/L less), particularly among specimens with high 25-OH vitamin D2 concentrations (average of 34.1 nmol/L less).
Biospecimens
FluidPlasmaPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
Small moleculeLC-MS or LC-MS/MSSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level <2.5 mol/L 25-OH vitamin D2>25 nmol/L 25-OH vitamin D2LC-MS or LC-MS/MS Specific Technology platform Liaison analyzer