Impact of a new refrigerator for the preservation of whole blood.

Kanamoto M, Shimada M, Utsunomiya T, Imura S, Morine Y, Ikemoto T, Mori H, Hanaoka J

Publication: J Surg Res (2012) • Vol 173 • Page 180-4PMID: 20950827View on PubMed

Purpose

The purpose of this paper was to determine the effects of using a Keep fresh refrigerator equipped with sensors to detect internal and surface specimen temperatures rather than a traditional refrigerator for specimen storage on whole blood ammonia, lactate, and adenosine triphosphate (ATP) levels as well as pH.

Conclusion

The rates of increase of ammonia and lactate levels and the rates of decrease in pH and ATP were lower in specimens in Keep fresh than traditional refrigerators. Significant differences in ammonia, lactate, pH and ATP levels between specimens in Keep fresh and tradition refrigerators were found after 7-9, 9, 5-13, and 5-15 weeks, respectively, but with the exception of ATP differences were transient.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storing whole blood for up to 15 weeks in Keep fresh refrigerators equipped with sensors to detect internal and surface specimen temperatures rather than traditional refrigerators for storage on whole blood ammonia, lactate, and ATP levels as well as pH. Blood was anticoagulated with citrate-phosphate-dextrose-adenine (CPDA). The Keep fresh refrigerator was programmed for rapid cooling (2 degrees C/h) of the specimen to 8 degrees C followed by cooling at 0.4 degrees C/h to 4 degrees C.

Summary of Findings

The rates of increase of ammonia and lactate were significantly lower in specimens in Keep fresh than traditional refrigerators (p<0.01, and p<0.05, respectively). While concentrations of ammonia and lactate were initially similar between specimens in the two different refrigerator types, after 7 and 9 weeks, and after just 9 weeks, ammonia and lactate levels, respectively, were transiently higher in specimens stored in traditional refrigerators than specimens stored in Keep fresh refrigerators (p<0.05, all), and with longer storage, these differences were no longer significant. Overall, pH decreased during storage of specimens, but the rate of decline was much greater in specimens stored in traditional refrigerators than Keep fresh refrigerators (p<0.001) and declined significantly by 5 weeks, while pH was relatively stable over 9 weeks in the Keep fresh refrigerators. Further, pH was significantly higher in specimens stored in Keep fresh refrigerators between weeks 5 and 13 than in traditional refrigerators (p<0.05). Although ATP levels decreased with storage in both refrigerator types, declines in ATP were minimal during the first 7 weeks in Keep fresh refrigerators such that after 5 weeks or more, ATP was higher in specimens stored in Keep fresh refrigerators than in traditional refrigerators (p<0.01).

  • Biospecimens

    FluidBlood
  • Preservation Types

    Other Preservative
  • Diagnoses

    Normal
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
    Small moleculeBioassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage conditions
    Ordinary refrigerator
    Temperature sensing refrigerator
    StorageStorage duration
    1 week
    11 weeks
    13 weeks
    15 weeks
    3 weeks
    5 weeks
    7 weeks
    9 weeks