Effects of tourniquet technique, order of draw, and sample storage on plasma fibrinogen.

Rosenson RS, Staffileno BA, Tangney CC

Publication: Clin Chem (1998) • Vol 44 • Page 688-90PMID: 9510890View on PubMed

Purpose

The purpose of this paper was to determine the effects of tourniquet pressure and duration, sequential collection, and refrigerated storage of plasma on fibrinogen levels.

Conclusion

There were no significant differences in fibrinogen concentrations between the first collected specimen and the average of the next 4 specimens. When low or high tourniquet pressure was applied, fibrinogen concentrations progressively increased with the duration of tourniquet application, but there were no significant differences between the tourniquet pressures if tourniquet time was similar. Storage of plasma at 4 degrees C for up to 48 hours did not affect fibrinogen concentrations.

Studies (1)

Purpose

The purpose of this study was to determine the effects of tourniquet pressure and duration, sequential collection, and refrigerated storage of plasma on fibrinogen levels. During collection, a sphygmomanometer was used to standardize tourniquet pressure with low pressure defined as 40 mmHg and high pressure defined as mean arterial pressure + 10 mmHg for each volunteer. Tourniquet times of 0-1 min, 2-3 min, and 4-5 min were used. All blood was collected into citrate tubes. The first-stick effect was analyzed by comparing fibrinogen levels from the first specimen with the average fibrinogen levels from the next 4 specimens.

Summary of Findings

There were no significant differences in fibrinogen concentrations between the first collected specimen and the average concentration of the next 4 specimens. When low tourniquet pressure was applied (40 mmHg), fibrinogen concentrations progressively increased with the duration of tourniquet application (p<0.00005), but there was no statistically significant difference between specimens obtained with 2-3 min tourniquet application and those obtained with 4-5 min tourniquet application. When high tourniquet pressure was applied (mean arterial pressure + 10 mmHg), fibrinogen concentrations also progressively increased with the duration of tourniquet application (p<0.00005), but there was no statistically significant difference between specimens obtained with 2-3 min tourniquet application and those obtained with 0-1 min tourniquet application. Changing the tourniquet pressure did not affect the average fibrinogen concentrations, as long as the tourniquet time was similar. Storage of plasma at 4 degrees C for up to 48 hours did not affect fibrinogen concentrations.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    GlycoproteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionMethod of fluid acquisition
    Tourniquet pressures compared
    Tourniquet times compared
    Biospecimen Aliquots and ComponentsAliquot sequential collection
    1st collection
    Not first collection
    StorageStorage duration
    24 h
    48 h
    6 h
    <10 min