Effects of tourniquet technique, order of draw, and sample storage on plasma fibrinogen.
Rosenson RS, Staffileno BA, Tangney CC
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of tourniquet pressure and duration, sequential collection, and refrigerated storage of plasma on fibrinogen levels. During collection, a sphygmomanometer was used to standardize tourniquet pressure with low pressure defined as 40 mmHg and high pressure defined as mean arterial pressure + 10 mmHg for each volunteer. Tourniquet times of 0-1 min, 2-3 min, and 4-5 min were used. All blood was collected into citrate tubes. The first-stick effect was analyzed by comparing fibrinogen levels from the first specimen with the average fibrinogen levels from the next 4 specimens.
Summary of Findings
There were no significant differences in fibrinogen concentrations between the first collected specimen and the average concentration of the next 4 specimens. When low tourniquet pressure was applied (40 mmHg), fibrinogen concentrations progressively increased with the duration of tourniquet application (p<0.00005), but there was no statistically significant difference between specimens obtained with 2-3 min tourniquet application and those obtained with 4-5 min tourniquet application. When high tourniquet pressure was applied (mean arterial pressure + 10 mmHg), fibrinogen concentrations also progressively increased with the duration of tourniquet application (p<0.00005), but there was no statistically significant difference between specimens obtained with 2-3 min tourniquet application and those obtained with 0-1 min tourniquet application. Changing the tourniquet pressure did not affect the average fibrinogen concentrations, as long as the tourniquet time was similar. Storage of plasma at 4 degrees C for up to 48 hours did not affect fibrinogen concentrations.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
GlycoproteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Tourniquet pressures comparedTourniquet times comparedBiospecimen Aliquots and Components Aliquot sequential collection 1st collectionNot first collectionStorage Storage duration 24 h48 h6 h<10 min