Comparison of hemolysis in blood samples collected using an automatic incision device and a manual lance.
Kazmierczak SC, Robertson AF, Briley KP
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of patient age, gestational age, and birth weight, and using a spring-loaded versus manual lance for the collection of blood from newborns by heel prick on plasma hemoglobin concentrations.
Summary of Findings
Gestational age, patient age, and birth weight did not influence the level of hemoglobin found in plasma specimens. However, the use of an automatic spring-loaded lance (Tenderfoot device) for heel prick blood collection from newborns significantly reduced mean plasma hemoglobin concentrations compared to the use of a manual handheld lance (Microlance) (p<0.001).
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Patient age 7 to 126 hours oldGestational age of 33 to 41 weeksPreaquisition Patient body mass index Birth weight of 1254 g to 5267 gBiospecimen Acquisition Method of fluid acquisition Manual lanceSpring-loaded automatic lance