Skip to main content

Serum and plasma preparation for potassium analysis: effects of anticoagulants, storage time and temperature before separation, and haemolysis.

BERNSTEIN RE

Publication: S Afr J Med Sci (1953) • Vol 18 • Page 99-104PMID: 13135665View on PubMed

Purpose

The purpose of this paper was to determine the effects of whole blood storage temperature and duration and anticoagulant on potassium levels, hematocrit, and hemolysis rates in plasma and serum.

Conclusion

The authors report that all changes in potassium levels within the first 6 h of whole blood storage at any of the tested temperatures were clinically insignificant, but the authors hypothesize that longer storage may lead to false diagnosis of hypopotassemia, particularly in specimens stored at 4-8 degrees C. Hemolysis rates increased with increasing storage temperature, but the authors report these changes contributed minimally to potassium levels. The trends were similar between plasma and serum, but serum values showed greater variability. The hematocrit was 2-5% lower and the hemolysis rates were higher among specimens anticoagulated with sodium citrate or lithium oxalate compared to serum specimens or plasma anticoagulated with heparin.

Studies (1)

Purpose

The purpose of this study was to determine the effects of whole blood storage temperature and duration and anticoagulant on potassium levels, hematocrit, and hemolysis rates in plasma and serum. Whole blood specimens intended for plasma collection were inverted every 30 minutes during storage.

Summary of Findings

The hematocrit was 2-5% lower in specimens anticoagulated with sodium citrate or lithium oxalate than in heparin plasma or serum. Further, the authors report higher hemolysis rates when specimens were anticoagulated with sodium citrate, lithium oxalate, or sodium ammonium oxalate, leading the authors to use only heparin for stability studies. After storage of whole blood for 3 h at 4-8 degrees C, significant increases in the resultant plasma and serum potassium levels were observed. During storage of whole blood at 18-25 degrees C, potassium levels in serum and plasma were stable for 3-9 h. In contrast, when whole blood was stored at 37 degrees C, potassium levels in plasma and serum decreased during the first 6 h before increasing. Increasing the storage temperature led to increasing hemolysis rates, but the authors show this only contributed minimally to the increase in potassium levels. The authors report that all changes within the first 6 h were clinically insignificant, but they hypothesize that longer storage may lead to false diagnosis of hypopotassemia. While the trends were similar between plasma and serum, serum values showed greater variability.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalFlame emission photometry
    Cell count/volumeSpectrophotometry
    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Heparin
    Lithium oxalate
    None
    Sodium ammonium oxalate
    Sodium citrate
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    1 h
    12 h
    15 h
    24 h
    3 h
    36 h
    48 h
    6 h
    9 h
    <10 min
    StorageStorage temperature
    18-25 degrees C
    37 degrees C
    4-8 degrees C