Evaluation of sample hemolysis in blood collected by S-Monovette using vacuum or aspiration mode.
Lippi G, Avanzini P, Musa R, Sandei F, Aloe R, Cervellin G
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood collection tube type and collecting blood by vacuum versus forced aspiration on cell-free hemoglobin, potassium, and LD levels in serum.
Summary of Findings
None of the specimens collected had clinically significant hemolysis, and there were no significant differences in potassium or LD levels between specimens collected by vacuum in S-Monovette tubes or in Vacutainers. However, potassium and LD levels were significantly lower in specimens collected in S-Monovette tubes by aspiration (syringe-style draw) than in specimens collected in S-Monovette or Vacutainer tubes by vacuum. Importantly, all of these significant differences, ranging on average from 2.1-2.7%, were less than the allowable total error for each analyte.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Cell count/volumeSpectrophotometryProteinClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Syringe drawVacuum tubeBiospecimen Acquisition Type of collection container/solution S-Monovette serum tube (no gel separator)Vacutainer serum tube (no gel separator)