Neutrophil gelatinase-associated lipocalin: comparison of the use of EDTA and heparin plasma.

Bakker AJ, Syperda H

Publication: Clin Chem Lab Med (2012) • Vol 50 • Page 1679-80PMID: 22962229View on PubMed

Purpose

The purpose of this paper was to determine the effects of using heparin plasma rather than EDTA-plasma for the measurement of neutrophil gelatinase-associated lipocalin (NGAL) by the particle enhanced turbidimetric immunoassay (PETIA) assay.

Conclusion

NGAL levels in EDTA-plasma and heparin plasma were very strongly correlated (R=0.9978), but levels of NGAL were slightly higher in heparin plasma than in EDTA-plasma.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using heparin plasma rather than EDTA-plasma for the measurement of NGAL by the PETIA assay.

Summary of Findings

NGAL levels in EDTA-plasma and heparin plasma were very strongly correlated (R=0.9978). Levels of NGAL were slightly but significantly higher in heparin plasma than in EDTA-plasma with a constant bias of 10.4 ug/L. The authors conclude that either anticoagulant could be used for the measurement of NGAL by the PETIA assay.