Which point-of-care creatinine analyser for radiology: direct comparison of the i-Stat and StatStrip creatinine methods with different sample types.

Dimeski G, Tilley V, Jones BW, Brown NN

Publication: Ann Clin Biochem (2013) • Vol 50 • Page 47-52PMID: 23108764View on PubMed

Purpose

The purpose of this paper was to determine the effects of venous versus capillary collection, analysis of lithium heparin whole blood versus plasma, and whole blood collection container type on the measurement of creatinine by three different analyzers. In addition, interference in the measurement of creatinine was evaluated by spiking plasma with dopamine or dobutamine.

Conclusion

Creatinine measurements taken by the i-Stat analyzer were more highly correlated with the standard lithium heparin plasma Beckman DxC800 analyzer measurements than the StatStrip measurements, regardless of the type of specimen or collection container. Dopamine spiked into plasma caused analytically significant increases in the measurement of plasma creatinine by the i-Stat, but not when measured by the Beckman DxC800 or StatStrip analyzers. Higher concentrations of dobutamine caused analytically significant decreases in the measurement of plasma creatinine by the Nova StatStrip, but not the other two analyzers.

Studies (2)

Purpose

The purpose of this study was to determine the effects of venous versus capillary collection, analysis of lithium heparin whole blood versus plasma, and whole blood collection container type on the measurement of creatinine by three different analyzers.

Summary of Findings

For whole blood collected in a blood gas syringe, whole blood from a lithium heparin Vacuette, and lithium heparin plasma, the creatinine levels measured by the i-Stat analyzer correlated more strongly with the lithium heparin plasma measurements taken by the Beckman DxC800 analyzer (the standard) than the StatStrip measurements. Capillary whole blood (measured only by the StatStrip method) showed much greater differences in creatinine measurements with the standard than the three types of specimens measured by the i-Stat analyzer.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
    Other diagnoses
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Renal disease
    Biospecimen AcquisitionAnatomical location of blood draw
    Capillary
    Vein
    Biospecimen AcquisitionType of collection container/solution
    Blood gas syringe
    Vacuette lithium heparin tube
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Whole blood
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Beckman DxC800
    StatStrip
    i-Stat