A comparison of glucose concentration in paired specimens collected in serum separator and fluoride/potassium oxalate blood collection tubes under survey 'field' conditions.

Fernandez L, Jee P, Klein MJ, Fischer P, Perkins SL, Brooks SP

Publication: Clin Biochem (2013) • Vol 46 • Page 285-8PMID: 23219741View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type, patient gender, and fasting status on glucose measurements and to evaluate the effects of tube type on hemolysis levels in plasma and serum.

Conclusion

Although a much higher rate of hemolysis was observed for sodium fluoride/potassium oxalate (NaF/KOx) plasma specimens than for serum specimens collected in serum separator tubes (SST), neither tube type, gender, nor fasting status of the patient during blood draw significantly affected glucose levels.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube type, patient gender, and fasting status on glucose measurements and to evaluate the effects of collecting NaF/KOx plasma versus serum in an SST on hemolysis levels in plasma and serum. Draw order was dependent on whether the patient had been fasting or not fasting. For each patient, 4 specimens were drawn into SST and 1 specimen was collected into a NaF/KOx tube. Specimens were centrifuged immediately after mixing/clotting, or were placed at 4 degrees C for <2h before centrifugation. Plasma was aliquoted, stored at -20 degrees C, and shipped frozen on dry ice for analysis within 7 days of collection. Pooled serum (from the 4 draws) was aliquoted and shipped at 4-8 degrees C on ice packs for analysis within 7 days of collection.

Summary of Findings

While 86.2% of NaF/KOx tube specimens had hemolysis values >0.15 g/L, only 2.2% of SST specimens had hemolysis values that high. Only non-hemolyzed specimen pairs were used for further analysis. There were no significant differences in glucose levels between serum and plasma specimens, between specimens drawn from male or female patients, or between specimens from fasting and non-fasting individuals. Further, when specimen pairs were included in which the NaF/KOx plasma had a hemolysis value >0.15 g/L (initially excluded), there were still no significant differences in glucose levels between specimens collected in the two different tube types.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    FrozenOther Preservative
  • Diagnoses

    Not specified
  • Platforms

    Cell count/volumeSpectrophotometry
    CarbohydrateClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPatient diet
    Fasting
    Non-fasting
    PreaquisitionPatient gender
    Female
    Male
    Biospecimen AcquisitionType of collection container/solution
    NaF/KOx tube
    SST
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen PreservationType of fixation/preservation
    Frozen
    Refrigeration