Purpose
The purpose of this paper was to determine the effects of shipping specimens overnight to a central laboratory on complete blood count (CBC) results.
Conclusion
Local laboratory values for absolute neutrophil count (ANC) and absolute monocyte count (AMC) were higher, and absolute lymphocyte count (ALC) and mean corpuscular volume (MCV) were lower than those obtained after shipping specimens overnight to the central laboratory (p<0.0001, all). Further, the authors report the correlations between local and central values for ANC and AMC were very weak.
Studies (1)
Purpose
The purpose of this study was to determine the effects of shipping specimens overnight to a central laboratory on CBC results in infants with sickle cell anemia. Two aliquots of potassium EDTA blood were collected from 167 individuals. One specimen was analyzed in the local lab while the other was shipped to the central laboratory. Specimens were shipped overnight with frozen ice packs and inserts to keep the tube upright.
Summary of Findings
Local laboratory values for ANC and AMC were higher, and ALC and MCV were lower than those obtained after shipping specimens overnight to the central laboratory (p<0.0001, all). Further, the authors report the correlations between local and central values for ANC and AMC were very weak, while the correlations between local and central laboratory values for ALC and MCV were better, but the exact values were not given.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Other diagnosesPlatforms
Cell count/volumeHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Sickle cell anemiaStorage Between site transportation method MailedNot transported