Effects of vigorous mixing of blood vacuum tubes on laboratory test results.

Lima-Oliveira G, Lippi G, Salvagno GL, Montagnana M, Gelati M, Volanski W, Boritiza KC, Picheth G, Guidi GC

Publication: Clin Biochem (2013) • Vol 46 • Page 250-4PMID: 23146916View on PubMed

Purpose

The purpose of this paper was to determine the effects of shaking tubes vigorously versus gently inverting them on 13 hematological, 3 coagulation, and 32 clinical chemistry analytes in blood.

Conclusion

Vigorous shaking of tubes for 3-5 sec, rather than gentle inversion, did not affect the levels of the hematological, coagulation, or clinical chemistry analytes.

Studies (1)

Purpose

The purpose of this paper was to determine the effects of shaking tubes vigorously versus gently inverting them on 13 hematological, 3 coagulation, and 32 clinical chemistry analytes in blood. Blood was collected from 100 fasting volunteers, who remained seated for 15 min without venous stasis, into K2 EDTA vacuum tubes (hematological analytes), serum tubes with clot activator and gel separator (clinical chemistry analytes) and sodium citrate tubes (coagulation analytes). The first 2 mL of blood were discarded. Tubes were mixed immediately after collection, and plasma tubes were immediately centrifuged, while serum tubes were allowed to clot for 45 min before centrifugation. Hematological analysis was performed on fresh blood, but plasma and serum were stored at -70 degrees C until coagulation testing and clinical chemistry analysis, respectively.

Summary of Findings

Vigorous shaking of tubes for 3-5 sec, rather than gentle inversion, did not affect the levels of the hematological, coagulation, or clinical chemistry analytes. After shaking vigorously, foam was present in all specimens, and serum had a blood ring where the stopper had been.

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)Frozen
  • Diagnoses

    Normal
  • Platforms

    Cell count/volumeHematology/ auto analyzer
    ProteinHematology/ auto analyzer
    ProteinClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    GlycoproteinHematology/ auto analyzer
    PeptideClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    3-5 sec vigorous shaking
    Inverted gently as recommended