Impact of prandial status on the comparison of capillary glucose meter and venous plasma glucose measurements in healthy volunteers.

Swaminathan A, Lunt H, Chang WS, Logan FJ, Frampton CM, Florkowski CM

Publication: Ann Clin Biochem (2013) • Vol 50 • Page 6-12PMID: 23148281View on PubMed

Purpose

The purpose of this paper was to determine the effects of venous versus capillary blood collection, fasting versus non-fasting collection, gender, and type of glucometer on glucose measurements.

Conclusion

Venous plasma glucose levels were lower in specimens collected 1 hour after breakfast compared to those collected after an overnight fast. Overall, capillary specimens contained a higher average glucose concentration than venous plasma when either the FreeStyle Lite meter or the Performa meter was used. But for both meters, capillary blood glucose concentrations were lower than venous plasma concentrations after an overnight fast and were higher than venous plasma concentrations in specimens collected 1 hour after breakfast.

Studies (1)

Purpose

The purpose of this study was to determine the effects of venous versus capillary blood collection, fasting versus non-fasting collection, gender, and type of glucometer on glucose measurements in specimens from healthy volunteers (not diabetic). Following capillary collection, venous specimens were collected from the same arm, centrifuged immediately, and plasma was analyzed.

Summary of Findings

Venous plasma glucose levels were lower in specimens collected 1 hour after breakfast compared to those collected after an overnight fast by an average of 0.59 mmol/L. Neither the amount of carbohydrates consumed during breakfast nor gender influenced the change in glucose levels. Overall, capillary blood contained an average glucose concentration that was 0.15 mmol/L higher than venous plasma when the FreeStyle Lite meter was used and 0.53 mmol/L higher when the Performa meter was used. When the FreeStyle Lite meter was used, the average preprandial difference between capillary and venous specimen glucose concentration was -0.52 mmol/L (venous specimens were higher), and the average postprandial difference between capillary and venous specimens was 0.81 mmol/L (capillary specimens were higher). When the Performa meter was used, the average preprandial difference was -0.13 mmol/L (venous specimens were higher), and the average postprandial difference was 1.19 mmol/L (capillary specimens were higher). The differences between capillary and venous specimen glucose concentrations in the pre- and postprandial states are significant for both meters.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPatient diet
    24.2-296.8 g carbohydrates
    PreaquisitionPatient gender
    Female
    Male
    Biospecimen AcquisitionAnatomical location of blood draw
    Capillary
    Vein
    Biospecimen AcquisitionTime of biospecimen collection
    1 hour after breakfast
    After overnight fast
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Whole blood
    Clinical chemistry/auto analyzer SpecificTechnology platform
    FreeStyle Lite
    Performa