The effects of a freeze-thaw cycle and pre-analytical storage temperature on the stability of insulin-like growth factor-I and pro-collagen type III N-terminal propeptide concentrations: Implications for the detection of growth hormone misuse in athletes.
Guha N, Erotokritou-Mulligan I, Bartlett C, Cowan DA, Bassett EE, Stow M, Sönksen PH, Holt RI
Purpose
The purpose of this paper was to determine the effects of storage, a single freeze-thaw cycle, and immunoassay type on measurement of Insulin-like growth factor-I (IGF-I) and pro-collagen type III N-terminal propeptide (P-III-NP) in serum.
Conclusion
IGF-I and P-III-NP levels were unaffected by storage of serum for 24 h at 4 degrees C followed by freezing and storage at -20 degrees C for up to 87 days. Further, the growth hormone (GH)-2000 score calculated from the IGF-I and P-III-NP levels showed low variability, and the authors state the false-positive rate for performance enhancing drug abuse would be 1:10,000 for the time-points investigated. The intra-specimen and inter-assay coefficients of variance (CVs) in IGF-I measurements were lower using the immunoradiometric assay (IRMA) than then ELISA. The P-III-NP intra-specimen CV was higher, but the inter-assay CV was lower using the Orion radioimmunoassay (RIA) than the Cisbio RIA.
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage, a single freeze-thaw cycle, and immunoassay type on measurement of IGF-I and P-III-NP in serum. Specimens were either analyzed immediately, transferred to 4 degrees C for 24 h, analyzed again and then transferred to -20 degrees C for the remainder of the storage.
Summary of Findings
IGF-I and P-III-NP levels were unaffected by storage of serum for 24 h at 4 degrees C followed by freezing and storage at -20 degrees C for up to 87 days. Further, the GH-2000 score calculated from the IGF-I and P-III-NP levels showed low variability, and the authors state the false-positive rate for performance enhancing drug abuse would be 1:10,000 for the time-points investigated. The intra-specimen and inter-assay CVs in IGF-I measurements were lower using the IRMA than then ELISA. The P-III-NP intra-specimen CV was higher, but the inter-assay CV was lower using the Orion RIA than the Cisbio RIA.
Biospecimens
FluidSerumPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
ProteinELISAProteinImmunoradiometric assayPeptideRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Freeze/thaw cycling 0 cycles1 cycleStorage Storage duration 0 days1 day2 days36 days8 days87 daysRadioimmunoassay Specific Technology platform Cisbio RIAOrion RIAELISA Specific Technology platform IRMA