Pre-analytical and analytical considerations for the determination of plasma renin activity.
Brossaud J, Corcuff JB
Purpose
The purpose of this paper was to determine the effects of blood storage temperature and assay duration on plasma renin activity (PRA) as determined by changes in angiotensin I concentrations.
Conclusion
When blood was stored for 5 h at 25 degrees C, PRA was significantly lower than levels measured in specimens stored at 4 degrees C for the same duration, but this decrease was attributed to an increase in the activity of the blank. The sensitivity and the precision of the PRA assay increased when the incubation time was increased from 1.5 to 18 h.
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood storage temperature and assay duration on PRA as determined by changes in angiotensin I concentrations. The blank used in the PRA assay contained the concentration of angotensin I in specimens assayed at 4 at the relevant reaction time instead of 37 degrees C. All specimens were frozen before analysis.
Summary of Findings
When blood from 14 subjects was stored for 5 h at 25 degrees C, PRA was significantly lower than levels measured in specimens stored for 5 h at 4 degrees C (p<0.002). However, this decrease could be attributed to a more than 50% increase in the activity of the blank (4 degree C reaction condition) and not a change in the angiotensin I concentration (37 degree C reaction condition). The sensitivity and the precision of the PRA assay increased when the incubation time was increased from 1.5 to 18 h.
Biospecimens
FluidBloodPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
ProteinEnzyme assayProteinRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage temperature 25 degrees C4 degrees CEnzyme assay Specific Incubation time 1.5 h18 h