Automatic Buffy Coat Extraction: Methodology, Feasibility, Optimization and Validation Study

Mathay C, Ammerlaan W, Betsou F

Publication: Biopres and Biobank (2012) • Vol 10 • Page 543-547

Purpose

The purpose of this paper was to determine the effects of manual rather than automated removal of buffy coat (BC) and blood volume on cell counts, DNA yield and DNA purity.

Conclusion

After automated BC removal, the percentages of lymphocytes and granulocytes were within the reference ranges for all three subjects, but the percentages of monocytes and CD19 B-cells were above the reference ranges, and CD3 and CD4 T cell percentages were below the reference ranges for 1 or 2 of the 3 donors. Automatically removed BC retained more of the leukocytes and platelets in the whole blood than manually prepared BC, but red blood cell percentages and hemoglobin content were not affected by removal method. While significantly less DNA was obtained from automatically removed BC rather than manually removed BC when 4 mL of blood was used (p=0.029), there was a trend to higher DNA yields from automatically removed BC than manually removed BC when 8 mL of blood was used. The DNA purity, as determined by spectrophotometric ratios, was unaffected by BC removal method.

Studies (1)

Purpose

The purpose of this study was to determine the effects of manual rather than automated BC removal and blood volume on cell counts, DNA yield and DNA purity. K2EDTA blood was centrifuged, and the BC was removed manually, or using a pipetting robot. DNA was extracted using an automated magnetic bead protocol.

Summary of Findings

After automated BC removal, the percentages of lymphocytes and granulocytes were within the reference ranges for all three subjects. In contrast, after automated BC removal, one of the three donors had monocyte and CD19 B cell percentages above the reference ranges and CD3 T cell percentages below the reference range. Another donor had CD3 and CD4 T cell percentages below the reference ranges after automated BC removal. For the third donor, all leukocyte subpopulation percentages remained within the reference ranges after automated BC removal. Automatically removed BC retained a higher percentage of leukocytes (95% versus 56%, p=0.008) and platelets (69% versus 35%, p=0.043) in the whole blood than manually prepared BC, but red blood cell percentages and hemoglobin content were not affected by BC removal method. While significantly less DNA was obtained from automatically removed BC rather than manually removed BC when 4 mL of blood was used (p=0.029), there was a trend to higher DNA yield from automatically removed BC than manually removed BC when 8 mL of blood was used. The DNA purity, as determined by spectrophotometric ratios, was unaffected by BC removal method.

  • Biospecimens

    FluidBlood
    FluidBuffy Coat
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Cell count/volumeHematology/ auto analyzer
    DNASpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    4 mL
    8 mL
    Biospecimen Aliquots and ComponentsBlood processing method
    Automated buffy coat removal
    Manual buffy coat removal