Lack of consistency between two commercial ELISAs and against an in-house ELISA for the detection of CD36 in human plasma.

Lykkeboe S, Larsen AL, Handberg A

Publication: Clin Chem Lab Med (2012) • Vol 50 • Page 1071-4PMID: 22706248View on PubMed

Purpose

The purpose of this paper was to determine the effects of centrifugation speed, freeze-thaw cycling, and ELISA manufacturer on the quantification of thrombospondin receptor (CD36) in EDTA-plasma.

Conclusion

CD36 levels, as measured by each of the three assays, were strongly or very strongly correlated between specimens centrifuged at 1000 x g and those centrifuged at 1850 x g. Further, CD36 levels were strongly correlated between fresh specimens and those subjected to 3 freeze-thaw cycles for each of the three assays. However, CD36 levels were not correlated between the assays, and levels differed by 2 to more than 1000 fold between ELISA methods.

Studies (1)

Purpose

The purpose of this study was to determine the effects of centrifugation speed (1000 x g versus 1850 x g), freeze-thaw cycling, and ELISA manufacturer on the quantification of CD36 in EDTA-plasma. The in-house protocol required 3 freeze-thaw cycles.

Summary of Findings

CD36 levels in specimens centrifuged at 1000 x g and 1850 x g were very strongly correlated when measured by the Adipobioscience ELISA (r=0.99, p<0.001) and strongly correlated using the Cusabio Biotech (r=0.80, p=0.006) and in-house ELISAs (r=0.85, p=0.002). Further, CD36 levels in specimens subjected to 3 freeze-thaw cycles (in house ELISA) and those subjected to none were very strongly correlated when measured by the Adipobioscience ELISA (r=0.91, p<0.001) and strongly correlated using the Cusabio Biotech ELISA (r=0.81, p=0.004). However, CD36 levels determined by the three different ELISAs were not correlated with each other and differed by 2 to more than 1000 fold. Cusabio Biotech ELISA values were at the low end of the assay sensitivity making dilution not possible. The Adipobioscience ELISA had a higher than optimal background.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    ProteinELISA
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Multiple speeds compared
    StorageFreeze/thaw cycling
    0 cycles
    3 cycles
    ELISA SpecificTechnology platform
    Adipobioscience
    Cusabio Biotech
    In house