Effects of sample transportation on commonly requested laboratory tests.

Zaninotto M, Tasinato A, Padoan A, Vecchiato G, Pinato A, Sciacovelli L, Plebani M

Publication: Clin Chem Lab Med (2012) • Vol 50 • Page 1755-60PMID: 23089704View on PubMed

Purpose

The purpose of this paper was to determine the effects of using an integrated transport system with a secondary and tertiary container and a data-logger on levels of alanine aminotransferase (ALT), prostate specific antigen (PSA), hemoglobin (Hb), calcium and potassium in blood.

Conclusion

Blood levels of potassium and ALT were affected by transport durations of more than 1.5 h when transported without the integrated transport system, but these parameters were not affected by transport duration after establishment of the integrated transport system. PSA, calcium and Hb were unaffected by transport duration before or after establishment of the integrated transport system.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using an integrated transport system with a secondary and tertiary container and a data-logger on the levels of ALT, PSA, Hb, calcium and potassium in blood. Specimens transported in the integrated transport system were rejected if temperatures did not remain between 20 and 25 degrees C and if transport required more than 3 h.

Summary of Findings

Potassium and ALT were affected by transport of blood for durations lasting more than 1.5 h when specimens were transported without the integrated transport system during at least 2 two month collection periods. However, when specimens were transported using the integrated transport system, potassium, and ALT were not significantly affected by transport duration, with the exception of potassium in specimens collected in November-December and transported for 1-1.5 h compared to those transported for <1 h (p=0.006). Levels of PSA, calcium and Hb were not affected by transport duration before or after establishment of the integrated transport system.

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    MorphologyHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionTime of biospecimen collection
    January-February
    July-August
    March-April
    May-June
    November-December
    September-October
    StorageBetween site transportation method
    Courier
    StorageSpecimen transport duration/condition
    1-1.5 h
    <1 h
    >1.5 h
    Integrated transport system
    No integrated transport system