The purpose of this study was to determine the effects of real and simulated transport of plasma and serum on potassium levels. Courier transport took 30 min, but the duration of simulated transport was unspecified.
Summary of Findings
Serum had on average 0.30 mmol/L higher potassium concentration than plasma, but this difference was insignificant. When plasma was subjected to simulated transport (jostling of tube) and an unspecified storage duration, average potassium levels increased by 0.20 mmol/L (p<0.0001) compared to initial levels. There was no difference between potassium levels in PST and SST specimens after courier transport. The authors also found lower potassium levels when specimens were transferred to a new tube prior to transport than when transported in the original tube, but the tube types used was unclear. The authors conclude that SST are the appropriate tube for potassium analysis at satellite sites.
Biospecimens
FluidPlasmaFluidSerumPreservation Types
None (Fresh)Diagnoses
NormalNot specifiedPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products PlasmaSerumStorage Between site transportation method CourierNot transportedStorage Specimen transport duration/condition In new PSTIn original PSTStorage Storage conditions Jostling upright to simulate the transportPrior to jostling,