Effect of acceleration forces during transport through a pneumatic tube system on ROTEM® analysis.
Amann G, Zehntner C, Marti F, Colucci G
Purpose
The purpose of this paper was to determine the effects of pneumatic tube system (PTS) acceleration on coagulation parameters in sodium citrate blood analyzed with ROTEM.
Conclusion
All ROTEM results for the hand-delivered and PTS specimens were within the normal range, but significant differences between coagulation parameters were observed between PTS and hand-delivered specimens.
Studies (1)
Purpose
The purpose of this study was to determine the effects of PTS acceleration on coagulation parameters in sodium citrate blood analyzed with ROTEM. Median transport times were 157 s for hand-delivery (A), 135 s for PTS at 3 m/s (B), and 176 s, 334 s, and 652 s for samples transported by PTS at 7 m/s for 270 m (C), 540 m (D) and 1080 m (E), respectively. Average times from collection until analysis for samples A, B, C, D and E were 11 min, 17 min, 81 min, 130 min and 157 min, respectively.
Summary of Findings
All ROTEM results for the hand-delivered and PTS specimens were within the normal range, but specimens delivered by PTS at 3 m/s (B)displayed small but significant changes in extrinsic (EXTEM) clot thickness at 10 (A10; p=0.021), 15 (A15; p<0.001) and 30 min (A30; p<0.001), maximum clot firmness (MCF; p<0.001), and area under curve (AUC; p<0.001), intrinsic (INTEM) A15 (p=0.024), A30 (p=0.004), MCF (p=0.027) and AUC (p=0.033), and aprotinin inhibited (APTEM) A30 (p<0.001), MCF (p=0.021) and AUC (p=0.008) compared to hand-delivered specimens. While remaining within the normal range, specimens transported at 7 m/s showed all the same significant differences observed between specimens transported at 3 m/s and hand-delivered specimens as well as significant differences in clot thickness at 5 min (A5), clot formation time (CFT), clotting time (CT), and alpha angle compared to hand-delivered specimens, but the degree of difference was dependent on transport distance. Increased transport distance, and the corresponding increases in acceleration forces and delays in analysis, resulted in larger effects on INTEM, EXTEM and APTEM, but not fibrin-based (FIBTEM) parameters.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Specimen transport duration/condition PTS 270 m at 7 m/sPTS 1080 m at 7 m/sPTS 270 m at 3 m/sPTS 540 m at 7 m/sStorage Within hospital transportation method Hand-deliveredPneumatic tube system