Serum testosterone quantitation by liquid chromatography-tandem mass spectrometry: interference from blood collection tubes.

Shi RZ, van Rossum HH, Bowen RA

Publication: Clin Biochem (2012) • Vol 45 • Page 1706-9PMID: 22971570View on PubMed

Purpose

The purpose of this paper was to determine the effects of blood collection tube type on testosterone quantitation by liquid chromatography tandem mass spectrometry (LC-MS/MS).

Conclusion

Interference by a co-eluting compound was observed in the testosterone elution window when serum was collected in either Vacutainer serum separator tubes (SST) or Vacutainer plasma separator tubes (PST), but not when serum was collected in Vacutainer Rapid Serum Tubes (RST), Vacutainer serum tubes, or Vacuette Z Serum Separator Clot Activator tubes. The interference increased with the length of time serum was held in the tubes, and was more pronounced for specimens with low testosterone and at the 289.4> 97.1 mass transition as opposed to the 289.4> 109.1 mass transition. The authors state that altering MS or LC parameters did not eliminate the testosterone peak interference and that SST and PST should be avoided for LC-MS/MS analysis of testosterone.

Studies (1)

Purpose

The purpose of this study was to determine the effects of blood collection tube type on testosterone quantitation by LC-MS/MS. The authors do not give details about specimen collection other than tube type.

Summary of Findings

Interference by a co-eluting compound was observed in the testosterone elution window when serum was collected in either Vacutainer SST or Vacutainer PST, but not when serum was collected in Vacutainer RST, Vacutainer serum tubes, or Vacuette Z Serum Sep Clot Activator tubes. The interference increased with the length of time serum was held in the tubes, and was more pronounced for specimens with low testosterone and at the 289.4> 97.1 mass transition as opposed to the 289.4> 109.1 mass transition. After further investigation, it was determined that the separator gel in the SST and PST was the main cause of interference, with a minor contribution from the stoppers, and no contribution from tube coating. The authors state that after altering MS or LC parameters such as using different buffers with varying concentrations and gradients, longer run times, using atmospheric-pressure chemical ionization (APCI) rather than electrospray ionization (ESI), and varying the source temperature, parameters of source gas, voltage, or collision energy for the transitions, they were unable to prevent the testosterone peak interference and that SST and PST should be avoided for LC-MS/MS analysis of testosterone.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    SteroidLC-ESI-MS/MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    High testosterone
    Low testosterone
    No testosterone
    Biospecimen AcquisitionType of collection container/solution
    Vacuette Z Serum Sep Clot Activator
    Vacutainer PST
    Vacutainer RST
    Vacutainer SST
    Vacutainer serum tube
    StorageStorage duration
    0 h
    0.5 h
    1 h
    10 h
    2 h
    24 h
    4 h
    86 h