BD rapid serum tubes reduce false positive plasma troponin T results on the Roche Cobas e411 analyzer.

Koch CD, Wockenfus AM, Saenger AK, Jaffe AS, Karon BS

Publication: Clin Biochem (2012) • Vol 45 • Page 842-4PMID: 22561083View on PubMed

Purpose

The purpose of this paper was to determine the effects of collecting serum in Rapid Serum Tubes (RST) rather than plasma in Plasma Separator Tubes (PST) on troponin T measurements.

Conclusion

Clinically concordant troponin T levels were measured between the specimens collected in RST and those collected in PST in 433 of 443 cases (98%).

Studies (1)

Purpose

The purpose of this study was to determine the effects of collecting serum in RST rather than plasma in PST on troponin T measurements. For discordant results between specimens collected in RST and PST, 3 and 6 h serial troponin T results from specimens collected in PST were used to determine which baseline measurement was the outlier. Positive or detectable tropoinin T levels were greater than or equal to 0.01 ng/mL. Discordance was defined as troponin T levels that differed by more than 0.02 ng/mL if both specimens had detectable levels or when only one specimen had detectable levels.

Summary of Findings

Clinically concordant troponin T levels were measured between the specimens collected in RST and those collected in PST in 433 of 443 cases (98%). A very strong correlation (r2=0.9987) and a mean bias of 0.000 ng/mL were calculated. Of the 10 discordant results, 3 had troponin T concentrations >0.25 ng/mL in both specimen types, 4 were classified as false positive baseline PST specimens, as the RST specimens and 3 and 6 h PST specimens were all negative for troponin T, 2 cases were classified as false negative RST specimens, as the baseline, 3, and 6 h PST specimens were all positive for troponin T. The remaining discordant case was positive for troponin T in the baseline PST but negative in the RST, and no serial troponin T testing was done so a false positive/false negative classification could not be assigned.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Other diagnoses
  • Platforms

    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Suspected myocardial damage
    Biospecimen AcquisitionType of collection container/solution
    PST
    RST
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum