The new Roche Vitamin D Total assay: fit for its purpose?
Emmen JM, Wielders JP, Boer AK, van den Ouweland JM, Vader HL
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to compare the measurement of 25(OH)D in serum and plasma using the Roche Elecsys Vitamin D Total assay and to compare the results of this assay with LC-MS/MS, HPLC, and Liaison immunoassay measurements. Left-over serum specimens from three different laboratories that had been frozen at -20 degrees C were used. Each laboratory compared the Elecsys assay to one other measurement method. The Modular Analytics E170 and Cobas e601 analyzers were used for the Elecsys assay in different laboratories.
Summary of Findings
Using the Elecsys assay, serum 25(OH)D levels were not significantly different from levels measured in heparinated plasma. 25(OH)D levels measured by the Elecsys assay and the Liaison immunoassay were strongly correlated (r=0.86), while those measured by the Elecsys assay and LC-MS/MS or HPLC were very strongly correlated (r=0.93 and r=0.91, respectively). The authors report that some individual specimens showed large between-method differences in 25(OH)D measurements.
Biospecimens
FluidSerumFluidPlasmaPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
Small moleculeClinical chemistry/auto analyzerSmall moleculeHPLCSmall moleculeLC-MS or LC-MS/MSSmall moleculeImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products PlasmaSerumClinical chemistry/auto analyzer Specific Technology platform Cobas e601HPLCImmunoassayLC MS/MSModular Analytics E170