The analytic impact of a reduced centrifugation step on chemistry and immunochemistry assays: an evaluation of the Modular Pre-Analytics.
Koenders MM, van Hurne ME, Glasmacher-Van Zijl M, van der Linde G, Westerhuis BW
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of centrifuging plasma or serum separately for 10 min at 1885 x g instead of in the analyzer for 5 min at 1885 x g on the levels of 22 clinical chemistry analytes. After centrifugation, all specimens were immediately placed into the analyzer for automated processing.
Summary of Findings
Use of the analyzer's 5 min centrifugation rather than a separate 10 min centrifugation before loading the specimens in the analyzer resulted in no statistical differences and very strong correlations in the levels of all analytes except the hemolysis index in plasma and serum and the lipemia index in plasma. The authors attribute the higher lipemia index after the 5 min centrifugation compared to the 10 min centrifugation (p<0.0001) to turbidity. The authors excluded the hemolysis index from correlation analysis because of too many values below the detection limit of the assay.
Biospecimens
FluidPlasmaFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerPeptideClinical chemistry/auto analyzerCarbohydrateClinical chemistry/auto analyzerGlycoproteinClinical chemistry/auto analyzerCell count/volumeSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Aliquots and Components Blood processing method Centrifuged in analyzerCentrifuged manuallyBiospecimen Aliquots and Components Centrifugation AutomatedManualMultiple durations compared