Preanalytic variables of thrombin generation: towards a standard procedure and validation of the method.
Loeffen R, Kleinegris MC, Loubele ST, Pluijmen PH, Fens D, van Oerle R, Ten Cate H, Spronk HM
Purpose
Conclusion
Studies (3)
Purpose
The purpose of this study was to determine the effects of blood collection method and tube type on hemolysis, ETP and peak height in plasma triggered with different amounts of TF. 21-gauge conventional and butterfly needles were used along with a 20-gauge intravenous catheter. All collection containers contained 3.2% citrate. After collection, plasma was separated by a two-step centrifugation protocol, frozen in liquid nitrogen, and stored at -80 degrees C until analysis.
Summary of Findings
Significantly more hemolysis was observed after centrifugation when specimens were collected through butterfly needles or intravenous catheters than when collected through a conventional straight needle, but there were only insignificant difference in ETP and thrombin peak height, regardless of TF trigger. Blood collection tube type affected thrombin generation triggered without TF or with 1 pM TF, but not 5 pM TF. The highest mean ETPs were observed in plasma from blood collected in glass Vacutainer tubes followed by plastic SCAT tubes, but the addition of CTI to SCAT tubes completely inhibited clot formation.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
FrozenDiagnoses
NormalPlatforms
MorphologyHematology/ auto analyzerMorphologySpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Butterfly deviceIV catheterNeedleVenipunctureBiospecimen Acquisition Type of collection container/solution Glass vacutainerPlastic MonovettePlastic SCATPlastic SCAT with corn trypsin inhibitorPlastic VenojectPlastic vacutainerHematology/ auto analyzer Specific Reaction solution 0 pM TF1 pM TF5 pM TF
Purpose
The purpose of this study was to determine the effects of discarding the first collection tube, delayed centrifugation, and using a one-step rather than two-step centrifugation on ETP and peak height in plasma collected in glass Vacutainer or plastic Monovette tubes and triggered with different concentrations TF. Blood was collected by standard venipuncture on alternating arms into the two types of collection tubes, each of which contained 3.2% citrate. After centrifugation, plasma was frozen in liquid nitrogen and stored at -80 degrees C until analysis.
Summary of Findings
When specimens were collected in glass Vacutainers, ETP and thrombin peak height triggered without TF or with 1 pM TF were significantly higher in the first compared to the second collection tube (p<0.0001, both), but this was not observed when blood was collected into plastic Monovette tubes or when triggered with 5 pM TF. When whole blood was stored in glass Vacutainer tubes, ETP and peak height were not affected by a 6 h delay before centrifugation, but when whole blood was stored in a plastic Monovette tube during such a delay, ETPs triggered without TF or with 1 pM TF were increased compared to those observed when specimens were centrifuged immediately (p=0.008), and the largest increases were observed when storage was at room temperature rather than 4 or 37 degrees C. Peak height was increased in specimens stored for 6 h, in plastic Monovette tubes, at 4 degrees C or room temperature, prior to centrifugation, and triggered without TF compared to peak heights observed for specimens that were centrifuged immediately (p=0.0092 and p=0.0005, respectively). ETP was higher when specimens in plastic Monovette tubes triggered without TF 0 or with 1 pM TF were centrifuged at 2,000 x g for 15 min instead of being centrifuged at 2,000 x g for 5 min and then 10,000 x g for 10 min (p=0.0005 and p=0.0256, respectively) but not triggered with 5 pM TF. A similar increase ETP after one-step rather than two-step centrifugation was observed in specimens collected in glass Vacutainer tubes and triggered with 1 pM TF (p=0.0466), but no significant effect of a one-step rather than two-step centrifugation was observed when glass Vacutainer specimens were triggered with 0 or 5 pM TF.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
FrozenDiagnoses
NormalPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Glass VacutainerPlastic Monovette,Biospecimen Aliquots and Components Aliquot sequential collection 1st collection2nd collectionBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedDifferent number of centrifugation steps comparedStorage Storage duration 0 h6 hStorage Storage temperature 37 degrees C4 degrees CRoom temperatureHematology/ auto analyzer Specific Reaction solution 0 pm TF1 pM TF5 pM TF
Purpose
Summary of Findings
Biospecimens
FluidPlasmaPreservation Types
FrozenDiagnoses
NormalPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 10 min20 min30 min40 min50 min60 minStorage Storage temperature 37 degrees C4 degrees CRoom temperatureHematology/ auto analyzer Specific Reaction solution 0 pM TF1 pM TF