No effect of anticoagulant on Hb A(1c) analysis by the IFCC reference procedure.
Biswas N, Raut N, Mathew RG, Sil A, Manohar V
Purpose
The purpose of this paper was to determine the effects of 5 different anticoagulants in vacutainers on the measurement of glycohemoglobin (Hb A1c) in fresh hemolysate.
Conclusion
Hb A1c percentages were not significantly different between blood specimens collected into Vacutainers containing sodium heparin, lithium heparin, sodium fluoride, sodium citrate, or EDTA.
Studies (1)
Purpose
The purpose of this study was to determine the effects of 5 different anticoagulants in vacutainers on the measurement of Hb A1c in fresh hemolysate.
Summary of Findings
Hb A1c percentages were not significantly different between blood specimens collected into Vacutainers containing sodium heparin, lithium heparin, sodium fluoride, sodium citrate, or EDTA.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinCoulter counterGlycoproteinLC-MS or LC-MS/MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level High Hb A1c levelLow Hb A1c levelMedium Hb A1c levelBiospecimen Acquisition Anticoagulant EDTALithium heparinSodium citrateSodium fluorideSodium heparin