Stability of 27 biochemistry analytes in storage at a range of temperatures after centrifugation.
Taylor EC, Sethi B
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage container, duration and temperature, and freeze-thaw cycling on the levels of 27 biochemical analytes in serum. Specimens were thawed at 37 degrees C for 5 min before analysis or refreezing. For each patient, 6 tubes of blood were collected of which, one specimen was analyzed fresh, another was stored in the SST at 2-10 degrees C, and serum from the other 4 tubes were transferred to new tubes and stored at either -80 or -20 degrees C or room temperature. After experimental storage, all specimens were stored frozen at -80 degrees C until batch analysis.
Summary of Findings
With the exception of uric acid, which was stable in serum stored at room temperature for <2 h, all 26 other analytes were stable in serum stored at room temperature for at least 120 h; however, a spike in cholesterol was observed after 2 h at room temperature, but the levels decreased to within the manufacturers precision limit by 4 h. None of the analytes were affected by storage of serum in a new tube or with an intact gel for 1 week at 2-10 degrees C, but potassium levels were above the precision limits after day 4 in the single serum specimen (of 10) stored in a SST that had evidence of the gel breaking down starting at day 4. Levels of all analytes except LD and ALT were stable in serum stored at -20 degrees C for at least 12 weeks, but LD and ALT levels were only stable in serum stored at -20 degrees C for 1 week and 2 weeks, respectively. ALT, LD, and uric acid were elevated in specimens that were frozen and thawed one or more times compared to specimens that were never frozen, but no other analytes were affected by up to 3 freeze-thaw cycles.
Biospecimens
FluidSerumPreservation Types
FrozenNone (Fresh)Other PreservativeDiagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerElectrolyte/MetalIon selective electrodeSteroidClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerGasIon selective electrodePre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Type of fixation/preservation FrozenNone (fresh)RefrigerationStorage Freeze/thaw cycling 0 cycles1 cycle2 cycles3 cyclesStorage Storage duration 0 h1 week12 weeks120 h2 h2 weeks24 h4 h4 weeks48 h72 h8 h8 weeks96 hStorage Storage temperature -20 degrees C-80 degrees C2-10 degrees CRoom temperatureStorage Type of storage container New tubeSST