Comparison of Bayer Advia Centaur immunoassay results obtained on samples collected in four different Becton Dickinson Vacutainer tubes.
Morovat A, James TS, Cox SD, Norris SG, Rees MC, Gales MA, Taylor RP
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood collection tube type on the results of 15 immunoassays, including ones for a variety of hormones as well as CA125, CA15-3, vitamin B12, and folate, performed by the Bayer Advia Centaur analyzer.
Summary of Findings
Results were comparable between specimens collected in plain plastic Vacutainers, old SST II Advance (discontinued in October 2004), new SST II Advance, and lithium-heparin plasma from PST II for TSH, FT4, FT3, PTH, E2, and prolactin. Serum collected in new SST II Advance showed slightly increased average FSH levels (+4.0%) and decreased average vitamin B12 levels (-10.0%) compared to serum collected in the plain plastic tubes. Serum collected in old SST II Advance showed increased average cortisol levels (+11.2%) and decreased average vitamin B12 levels (-16.6%) compared to specimens collected in plain plastic tubes. Plasma collected in PST II showed increased average LH (+8.0%), FSH (+7.9%), and folate levels (+37.7%), and decreased average cortisol (-34.3%), testosterone (-22.8%), and progesterone levels (-11.8%), and also exhibited positive absolute bias for vitamin B12 and CA15-3 compared to serum collected in plain plastic tubes.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
GlycoproteinClinical chemistry/auto analyzerPeptideClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution New SST II AdvanceOld SST II AdvancePST IIPlain plasticBiospecimen Aliquots and Components Blood and blood products PlasmaSerum