Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes.
van Berkel M, Scharnhorst V
Purpose
The purpose of this paper was to determine the effects of transferring blood into blood gas syringes containing electrolyte-balanced heparin on positively charged electrolytes in blood.
Conclusion
Compared to electrolyte concentrations measured in blood from non-heparinized tubes, clinically significant negative biases were observed in ionized calcium and sodium in specimens from preset and monovette blood gas syringes containing electrolyte-balanced heparin. On the other hand, specimens transferred into pico 50-2 syringes yielded only minimal negative biases in electrolyte concentrations compared to those measured in blood from non-heparinized tubes.
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood gas syringes containing electrolyte-balanced heparin on positively charged electrolytes in blood. Preset and pico 50-2 syringes contained 80 international units (IU) of electrolyte-balanced heparin while monovette syringes contained 50 IU. Specimens were collected in plain tubes first and then transferred into the blood gas syringes in a random order.
Summary of Findings
Compared to levels in blood from non-heparinized tubes, decreased levels of ionized calcium and sodium were statistically and clinically significant in specimens from preset and monovette syringes. In blood from pico 50-2 syringes, slightly decreased levels of sodium were statistically significant although clinically irrelevant, and ionized calcium levels were comparable to levels measured in blood from non-heparinized tubes. Potassium concentrations measured in specimens from all three syringes were significantly lower than those measured in non-heparinized blood, although none of the decreases were clinically relevant. The authors report that small differences in pH values among specimens in the 4 collection containers contributed to ionized calcium concentration differences. They also report that the order in which they transferred blood into the three different electrolyte-balanced heparin syringes and increasing fill volumes had no effect on the negative biases observed.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant HeparinMultiple concentrations evaluatedNoneBiospecimen Acquisition Type of collection container/solution Monovette (Sarstedt)Non-heparinized tubePico 50-2 (Radiometer)Preset (BD)