Increased levels of HIV RNA detected in samples with viral loads close to the detection limit collected in Plasma Preparation Tubes (PPT).
Griffith BP, Mayo DR
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of using PPT rather than standard EDTA tubes for collection and storage of plasma on the detection of HIV RNA over a range of values, including several specimens with very low viral loads. Specimens collected in standard EDTA tubes were centrifuged within 6 hours, and aliquoted plasma was stored and transported frozen. PPT plasma was stored and transported frozen in the original tube after centrifugation.
Summary of Findings
40% of the 112 specimen pairs had no detectable HIV RNA in the plasma collected in the standard EDTA tubes but low levels of HIV RNA detected in the plasma from the PPT. However, there were no instances of HIV RNA detection in EDTA-plasma and not in the PPT plasma. The discordance between tube types occurred more frequently when the ultrasensitive procedure for the Cobas Amplicor HIV-1 Monitor Test Version 1.5 was used rather than the standard procedure. It also occurred more frequently when the viral loads were closer to the lower limit of detection (50 copies/mL). Specimens with high viral loads showed less discrepancy between tube types.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
FrozenDiagnoses
AIDS/HIV-relatedPlatforms
RNAReal-time qRT-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level High HIV viral loadLow HIV viral loadBiospecimen Acquisition Type of collection container/solution EDTA VacutainerPPTStorage Type of storage container CryovialPPTReal-time qRT-PCR Specific Technology platform Standard assayUltrasensitive assay