Increased levels of HIV RNA detected in samples with viral loads close to the detection limit collected in Plasma Preparation Tubes (PPT).

Griffith BP, Mayo DR

Publication: J Clin Virol (2006) • Vol 35 • Page 197-200PMID: 16253548View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type on the detection of human immunodeficiency virus (HIV) RNA in plasma over a range of values.

Conclusion

40% of the 112 specimen pairs had no detectable HIV RNA in the plasma collected in standard EDTA tubes but low levels of HIV RNA detected in the plasma from plasma preparation tubes (PPT). However, there were no instances of HIV RNA detection in EDTA-plasma and not in the PPT plasma. The discordance between tube types occurred more frequently when the ultrasensitive procedure for the Cobas Amplicor HIV-1 Monitor Test Version 1.5 was used rather than the standard procedure. It also occurred more frequently when the viral loads were closer to the lower limit of detection (50 copies/mL). Specimens with high viral loads showed less discrepancy between tube types.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using PPT rather than standard EDTA tubes for collection and storage of plasma on the detection of HIV RNA over a range of values, including several specimens with very low viral loads. Specimens collected in standard EDTA tubes were centrifuged within 6 hours, and aliquoted plasma was stored and transported frozen. PPT plasma was stored and transported frozen in the original tube after centrifugation.

Summary of Findings

40% of the 112 specimen pairs had no detectable HIV RNA in the plasma collected in the standard EDTA tubes but low levels of HIV RNA detected in the plasma from the PPT. However, there were no instances of HIV RNA detection in EDTA-plasma and not in the PPT plasma. The discordance between tube types occurred more frequently when the ultrasensitive procedure for the Cobas Amplicor HIV-1 Monitor Test Version 1.5 was used rather than the standard procedure. It also occurred more frequently when the viral loads were closer to the lower limit of detection (50 copies/mL). Specimens with high viral loads showed less discrepancy between tube types.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    AIDS/HIV-related
  • Platforms

    RNAReal-time qRT-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    High HIV viral load
    Low HIV viral load
    Biospecimen AcquisitionType of collection container/solution
    EDTA Vacutainer
    PPT
    StorageType of storage container
    Cryovial
    PPT
    Real-time qRT-PCR SpecificTechnology platform
    Standard assay
    Ultrasensitive assay