Platelet counts in capillary blood.
Feusner JH, Behrens JA, Detter JC, Cullen TC
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of using capillary versus venous blood, capillary tubes versus EDTA Vacutainers, specimen volume, and counting method on the determination of platelet counts. Capillary tubes contained both heparin and EDTA while Vacutainers contained K3EDTA and potassium sorbate.
Summary of Findings
For healthy individuals and thrombocytopentic patients, platelet counts using plasma from capillary specimens were significantly lower than platelet counts from venous specimens (p<0.001 and p=0.01, respectively). Venous blood platelet counts were not significantly different when specimens were transferred to capillary tubes or Vacutainers, suggesting it is the collection location rather than the tube which affects platelet counts. In all cases, the automated count agreed with the smear estimate of the platelet count. When 1.0 or 0.5 mL venous specimen volumes were used as opposed to 3.0 mL, average platelet counts were 5 or 4% lower (p=0.026 and p=0.055, respectively).
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalOther diagnosesPlatforms
Cell count/volumeCoulter counterCell count/volumeLight microscopyPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition ThrombocytopeniaBiospecimen Acquisition Anatomical location of blood draw CapillaryVeinBiospecimen Acquisition Anticoagulant EDTAHeparinPotassium EDTABiospecimen Acquisition Type of collection container/solution Capillary tubesVacutainer tubesBiospecimen Aliquots and Components Aliquot size/volume 0.5 mL1.0 mL3.0 mLCoulter counter Specific Technology platform Light Microscopy