False reduction in serum methadone concentrations by BD Vacutainer serum separator tubes (SSTTM).

Berk SI, Litwin AH, Du Y, Cruikshank G, Gourevitch MN, Arnsten JH

Publication: Clin Chem (2006) • Vol 52 • Page 1972-4PMID: 16998122View on PubMed

Purpose

The purpose of this paper was to determine the effects of taking pegylated interferon alpha-2b and collection tube type on serum methadone concentrations in specimens from patients in a methadone maintenance program.

Conclusion

The use of glass or plastic serum separator tubes (SST) as opposed to plain glass tubes resulted in significant decreases in mean serum methadone concentrations. Pegylated interferon alpha-2b administration for two weeks prior to specimen collection partially attenuated the differences. Nevertheless, the authors recommend against use of SST for collection of blood for serum methadone measurement.

Studies (1)

Purpose

The purpose of this study was to determine the effects of taking pegylated interferon alpha-2b and collection tube type on serum methadone concentrations in specimens from 8 patients in a methadone maintenance program. Both prior to and after 2 weeks of pegylated interferon alpha-2b administration, sets of 3 specimens were consecutively drawn into plain glass, then SST glass, and finally SST Plus plastic tubes. Serum from the plain glass tubes was removed and frozen at -100 degrees C immediately after centrifugation whereas the serum from each type of SST was kept at 4 degrees C in the collection tube for 4 hours before being poured off and frozen at -100 degrees C.

Summary of Findings

The use of glass or plastic SST as opposed to plain glass tubes resulted in significant decreases in mean serum methadone concentrations of 22.4% and 25.6%, respectively. The mean serum methadone concentrations were also significantly different between specimens collected in glass SST and plastic SST (p=0.04). Pegylated interferon alpha-2b administration for two weeks prior to specimen collection partially attenuated the differences between serum methadone concentrations from plain glass tubes and those from either glass or plastic SST, but this was only significant for the difference between plain glass tubes and glass SST (p=0.03). The authors report that refrigeration of serum in both plastic and glass SST prior to freezing is not likely to be the cause of the lowered methadone concentrations compared to those in serum from plain glass tubes which was frozen immediately after separation.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    AIDS/HIV-related
    Hepatitis
  • Platforms

    Small moleculeHPLC
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionOther drugs
    None
    Pegylated interferon alpha-2b
    Biospecimen AcquisitionType of collection container/solution
    Plain glass
    SST Plus plastic
    SST glass
    StorageStorage duration
    0 h
    4 h