Methodological and preanalytical evaluation of a RAGE immunoassay.
Wittwer C, Lehner J, Fersching D, Siegele B, Stoetzer OJ, Holdenrieder S
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collecting plasma rather than serum, delayed centrifugation, storage, and freeze-thaw cycling on serum sRAGE levels.
Summary of Findings
EDTA-plasma, heparinated-plasma and citrated-plasma had 95.4%, 97% and 88.1% as much sRAGE as serum. There was no effect of storing serum for up to 7 days at 4 or 25 degrees C or of a 6 h or 24 h room temperature delay before centrifugation. Further, up to 3 freeze-thaw cycles of serum did not significantly affect sRAGE levels compared to serum thawed once. The intra-assay imprecision of measurement of sRAGE in serum was 5.6-11%.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
FrozenDiagnoses
NormalPlatforms
ProteinELISAPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant CitrateEDTAHeparinNoneBiospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Freeze/thaw cycling 1 cycle2 cycles3 cyclesStorage Storage duration 0 h24 h48 h6 h7 daysStorage Storage temperature 25 degrees C4 degrees CStorage Time at room temperature 0 h24 h6 h