Capillary versus venous bedside blood glucose estimations.

Boyd R, Leigh B, Stuart P

Publication: Emerg Med J (2005) • Vol 22 • Page 177-9PMID: 15735263View on PubMed

Purpose

The purpose of this paper was to determine the effects of using venous versus capillary blood and using a bedside blood glucometer versus a laboratory multichannel analyzer for the measurement of blood glucose levels.

Conclusion

Despite very strong correlations between all three glucose measurements, the average blood glucose measured in the laboratory using a venous specimen was significantly lower than the average capillary or venous blood glucose measured using a bedside glucometer. In addition, the average bedside venous blood glucose measurement was significantly higher than the average bedside capillary blood glucose measurement.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using venous versus capillary blood and using a bedside blood glucometer versus a laboratory multichannel analyzer for the measurement of blood glucose levels. Capillary and venous blood specimens were collected simultaneously and glucose was measured at bedside using each. Half of the venous specimen was sent to the laboratory in a lithium heparin tube for further whole blood glucose estimation.

Summary of Findings

Despite very strong correlations between all three glucose measurements, the average blood glucose measured in the laboratory on a venous specimen by the Dade-Behring Multichannel Analyzer (7.075 mmol/L) was significantly lower than the average capillary (7.66 mmol/L) or venous blood glucose (7.99 mmol/L) measured using a bedside Medisense Precision Plus Glucometer (both p<0.001). In addition, the average bedside venous blood glucose measurement was significantly higher than the average bedside capillary blood glucose measurement (p<0.05).

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnatomical location of blood draw
    Capillary
    Vein
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Dade-Behring Multichannel Analyzer
    Medisense Precision Plus Glucometer